Skip to the content

FDA Medical Device Recalls (openFDA)

Terumo BCT, Inc.: HARVEST(R) TERUMOBCT, BMAC-30-07 Bone Marrow Procedure Pack, Private Practice / Clinical Use Only, REF 51416, Rx Only, STERILE EO

Recall number
Z-0103-2018
Recalling firm
Terumo BCT, Inc.
Product
HARVEST(R) TERUMOBCT, BMAC-30-07 Bone Marrow Procedure Pack, Private Practice / Clinical Use Only, REF 51416, Rx Only, STERILE EO
Status
Terminated
Initiated
2017-10-20
Posted
not on file
Terminated
2020-08-21
Reason
The patient labels included in the BMAC system procedure packs, catalog numbers 51416 and 51423, are printed with an incorrect lot number, catalog number, and expiry date; the lot number on the procedure pack box label is correct. The expiration noted was not beyond initiation date of the recall. Since the product is noted as sterile, the clearance for K103340, K052925 and K991430 raises question if product was cleared with a sterility claim. Further post-market follow-up will be required to assess the safety risk of the product.
Root cause
Process control
Action
The firm initiated their recall by letter on 10/20/2017. The letter stated the following: "ACTIONS REQUIRED FOR HEALTHCARE PROVIDERS AND DISTRIBUTORS 1. For product shelved in the original procedure pack (white box), examine the label for catalog/lot numbers as shown in Figure C above. 2. If affected product is in your inventory, contact the Terumo BCT Customer Support Center at +1.877.3.FYI.BCT (U.S. Toll-Free +1.877.339.4228) or +1.303.231.HELP (+1.303.231.4357), Terumo BCT Europe N.V. at +32.2.715.0590, or your local Terumo BCT representative, and Terumo BCT will issue a Return Goods Authorization for product return. 3. Continue to use your unaffected product in accordance with the instructions for use. 4. Distribute this notification to all Harvest Bone Marrow Aspirate Concentrate (BMAC) System users within your organization. 5. IMPORTANT: Complete the attached acknowledgement and return it by fax or email to Terumo BCT by 27 October 2017. Your return of the acknowledgement is critical so we can confirm that you have received the recall notice.
Quantity
30 units
Distribution
CA, Co, MO, NY, PA, TX, VA, WI, Turkey, Austria, Spain
Lot, serial or model codes
Lot 04A9938
510(k) numbers
["K052925","K103340","K991430"]
PMA numbers
not on file
Product code
JQC
Firm FEI
1000117351
Firm city
Lakewood
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register