FDA Medical Device Recalls (openFDA)
Conmed Corporation: ConMed Laparoscopic Instrument: Meeker (Right Angle) Disposable Grasper, 5mm x 33cm length, REF/Product Code 60-6045-004. --- The product is distributed sterile in heat-sealed tray and labeled with an expiration date that pertains only to the sterility.
- Recall number
- Z-0103-2007
- Recalling firm
- Conmed Corporation
- Product
- ConMed Laparoscopic Instrument: Meeker (Right Angle) Disposable Grasper, 5mm x 33cm length, REF/Product Code 60-6045-004. --- The product is distributed sterile in heat-sealed tray and labeled with an expiration date that pertains only to the sterility.
- Status
- Terminated
- Initiated
- 2006-10-12
- Posted
- 2006-10-27
- Terminated
- 2012-08-30
- Reason
- The company was made aware of instances where the grasper jaws broke during laparoscopic procedures. When the grasper jaws had broken they had done so at the junction of the jaw and the tube. In no instances was it reported to the company that small pieces had broken off of the grasper jaws.
- Root cause
- Nonconforming Material/Component
- Action
- On 10/12/06, ConMed sent "URGENT DEVICE RECALL" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned.
- Quantity
- 38 units
- Distribution
- Worldwide
- Lot, serial or model codes
- REF/Product Code 60-6045-004, all lot codes manufactured between 08/25/2003 and 03/24/2005. [Lot codes on unopened boxes and packaging contain a lot code in the following form: for example 0409271 - 04 represents the year 2004, 09 represents the month of the year, 27 represents the day of the month, and the last digit 1 represents the manufacturing shift code. Lot codes are located on the device in this form: for example 40927 - 4 represents the year 2004, 09 represents the month of the year, and 27 represents the day of the month.]
- 510(k) numbers
- ["K943327"]
- PMA numbers
- not on file
- Product code
- GCJ
- Firm FEI
- 3001451571
- Firm city
- Utica
- Firm state
- NY
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28