Skip to the content

FDA Medical Device Recalls (openFDA)

OriGen Biomedical, Inc.: Reinforced Dual Lumen Cannula, 13Fr., REF: VV13F, Sterile EO; and ECMO Reinforced Dual Lumen Cannula, 13Fr., REF: VV13F, Sterile EO

Recall number
Z-0102-2019
Recalling firm
OriGen Biomedical, Inc.
Product
Reinforced Dual Lumen Cannula, 13Fr., REF: VV13F, Sterile EO; and ECMO Reinforced Dual Lumen Cannula, 13Fr., REF: VV13F, Sterile EO
Status
Terminated
Initiated
2018-08-21
Posted
not on file
Terminated
2021-10-05
Reason
The catheter failed the endotoxin testing.
Root cause
Release of Material/Component prior to receiving test results
Action
On 08/21/18, the firm emailed a recall notice to customers informing them of a nationwide recall of RDLC ECMO catheters. Customers were advised to return affected product to the firm. Customers with additional questions were advised to call 1-800-233-9014 or 1-512-474-7278. The firm also sent customers Technical Bulletin (TB) 1411 - Revision Feb 2018, Subject: Radiopacity of the Reinforced Dual Lumen Catheter (RDLC), Date: July 2015, and an Acknowledgment Receipt form to complete and return. On 09/12/18, and 09/20/18, the firm sent two additional recall notices, which included additional affected lot numbers that were not the subject of the current removal.
Quantity
106
Distribution
US Nationwide in the states of NY, AR, MO, NC, GA, LA, CO, AL, MA, CA, MI, TX, OH, MD, FL, TN, SC, OR, AZ, UT, RI, SD, PA, NM, CT. Foreign (OUS): United Arab Emirates, Kingdom of Saudi Arabia, New Zealand, Thailand
Lot, serial or model codes
UDI: 00816203022059, lot N18687-1
510(k) numbers
["K113869"]
PMA numbers
not on file
Product code
DWF
Firm FEI
1000160256
Firm city
Austin
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register