Siemens Medical Solutions USA, Inc: Syngo Imaging with versions VB20B, VB20D, VB20F, VB20G, VB25B, VB30A, VB30A_SP1, and VB30A_SP2.

Recall number
Z-0102-2009
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Syngo Imaging with versions VB20B, VB20D, VB20F, VB20G, VB25B, VB30A, VB30A_SP1, and VB30A_SP2.
Status
Terminated
Initiated
2008-08-22
Posted
2008-10-15
Terminated
2008-11-19
Reason
Image may not be visible after merge.
Root cause
Software design
Action
The recalling firm issued Customer Safety Advisory Notice dated 8/22/08 to inform the consignees of the potential issue and provide instruction to avoid its occurrence. To avoid potential risk of this issue, users are asked to check whether the extended HL7 messages as stated in the letter are mandatory for their RIS-PACS workflow, and disable these messages if they are not. Siemens is preparing a modification under the UI IM086/08/S for the affected versions of syngo Imaging that will resolve the potential malfunction. The field modification will be immediately distributed and installed by the firm's Service as soon as it is available.
Quantity
59 units
Distribution
Nationwide Distribution --- including states of AL, AZ, CA, CO, CT, FL, IA, KS, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NY, OH, OK, OR, PA, TN, TX, and WI.
Lot, serial or model codes
Model number 10014063. Serial numbers: 1005, 1039, 1040, 1041, 1049, 1053, 1055, 1056, 1057, 1058, 1061, 1064, 1065, 1067, 1071, 1074, 1076, 1079, 1080, 1081, 1082, 1083, 1084, 1086, 1089, 1090, 1095, 1105, 1107, 1112, 1116, 1119, 1124, 1174, 1188, 1198, 1202, 1203, 1210, 1218, 1226, 1227, 1229, 1240, 1246, 1247, 1258, 1269, 1271, 1278, 1279, 1282, 1316, 1322, 1329, 1334, 1343, 1351, and 1353.
510(k) numbers
["K052461","K071114"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register