Skip to the content

FDA Medical Device Recalls (openFDA)

Steris Corporation Hopkins Facility: Steris QFC1728 and QFC1729 Quick Connects;(used to connect gastrointestinal endoscopes to processing trays) The malfunctioning piece of the quick connect component is adaptor #6, STERIS part number 201158. Adaptor #6 attaches to the suction port of the endoscope.

Recall number
Z-0101-2008
Recalling firm
Steris Corporation Hopkins Facility
Product
Steris QFC1728 and QFC1729 Quick Connects;(used to connect gastrointestinal endoscopes to processing trays) The malfunctioning piece of the quick connect component is adaptor #6, STERIS part number 201158. Adaptor #6 attaches to the suction port of the endoscope.
Status
Terminated
Initiated
2007-09-12
Posted
2007-10-24
Terminated
2010-09-28
Reason
Loose or missing adaptor: The center insert of the adaptor #6 restrictor could become loose and/or come out of the adaptor's housing. The absence of adaptor #6's center insert may affect the sterilization process of the attached endoscope.
Root cause
Component design/selection
Action
STERIS has contacted affected direct accounts by phone to notify direct accounts of the recall. On 9/28/2007, STERIS will mail a recall notification to all affected direct accounts via certified mail. The notification informs direct accounts that STERIS representatives will be visiting direct accounts to retrieve affected products and provide replacements. Where direct accounts further distribute, STERIS representatives will inform the direct accounts to contact their sub-accounts and facilitate exchange of affected product for replacement product. Replacement product will be available beginning November 1, 2007.
Quantity
391 Quick Connects
Distribution
Worldwide: United States, Australia, Bahrain, Canada, Germany, Great Britain, Hungary, Italy, Qatar, and Spain
Lot, serial or model codes
C041782, C041793, 3072287, C041514, C041512, C041508, C041513, C041515, 3072303, 3290749, 3338126, C041781, C041794, 3072246, C041506, C041509, C041510, C041511, 3072261, 3290723, 3318292, 3338100, 3352374, 3542412, 3643871, 3546157, 3656121, 3893160, 3890027, 3748456, 3930062, 3978731, 3342342, 3352390, 3575941, 3643897, 3668357, 3895661, 3900198
510(k) numbers
not on file
PMA numbers
not on file
Product code
MED
Firm FEI
3000251274
Firm city
Mentor
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06