Zimmer Biomet, Inc.: ROSA Surgical Device 2.5.8 ROSA Surgical Device is a computer-controlled electromechanical arm. It is intended to be used in the operating room for the spatial positioning and orientation of an instrument holder or tool guide. Guidance is based on a pre-operative plan developed with three-dimensional imaging software, and uses fiducial markers or optical registration. The system is intended for use by neurosurgeons to guide standard neurosurgical instruments. It is indicated for any neurosurgical condition in which the use of stereotactic surgery may be appropriate
- Recall number
- Z-0100-2018
- Recalling firm
- Zimmer Biomet, Inc.
- Product
- ROSA Surgical Device 2.5.8 ROSA Surgical Device is a computer-controlled electromechanical arm. It is intended to be used in the operating room for the spatial positioning and orientation of an instrument holder or tool guide. Guidance is based on a pre-operative plan developed with three-dimensional imaging software, and uses fiducial markers or optical registration. The system is intended for use by neurosurgeons to guide standard neurosurgical instruments. It is indicated for any neurosurgical condition in which the use of stereotactic surgery may be appropriate
- Status
- Terminated
- Initiated
- 2012-04-19
- Posted
- not on file
- Terminated
- 2020-06-16
- Reason
- Potential failure of the optical distance sensor due to cable disconnection.
- Root cause
- Device Design
- Action
- Zimmer Biomet sent an Urgent Medical Device Recall Correction letter dated July 26, 2017, to all affected customers. Customers were informed on site of the device correction by Field Service Technicians. For questions regarding this recall call 574-267-6131.
- Quantity
- 14 units
- Distribution
- Worldwide Distribution - US including OH and MI Internationally to Canada, China, France, Germany, and Italy
- Lot, serial or model codes
- Serial No. RO10011 and RO10014
- 510(k) numbers
- ["K092239","K101791"]
- PMA numbers
- not on file
- Product code
- HAW
- Firm FEI
- 1000220733
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28