Sysmex America Inc: pocH-100i; sysmex; Automated Hematology Analyzer; 7.3" x 13.8" x 18.1" / 30.8 lbs.; Sysmex Corporation; 1-5-1 Wakinohama-Kaigandori, Chuo-ku, Kobe 651-0073, Japan; Sysmex America, Inc; One Nelson C. White Pkwy, Mundeline, IL 60060, U.S.A. The Sysmex pocH-100i Automated Hematology Analyzer is an automated cell counter intended for in vitro diagnostic use in CLIA non-waived clinical laboratories (not for Point of Care use in a CLIA waived laboratory). The Complete Blood Cel (CBC) count test parameters include: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, and NEUT%/#, LYMPH %/#, MIXED %/# (EO, BSO, MONO).

Recall number
Z-0100-2014
Recalling firm
Sysmex America Inc
Product
pocH-100i; sysmex; Automated Hematology Analyzer; 7.3" x 13.8" x 18.1" / 30.8 lbs.; Sysmex Corporation; 1-5-1 Wakinohama-Kaigandori, Chuo-ku, Kobe 651-0073, Japan; Sysmex America, Inc; One Nelson C. White Pkwy, Mundeline, IL 60060, U.S.A. The Sysmex pocH-100i Automated Hematology Analyzer is an automated cell counter intended for in vitro diagnostic use in CLIA non-waived clinical laboratories (not for Point of Care use in a CLIA waived laboratory). The Complete Blood Cel (CBC) count test parameters include: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, and NEUT%/#, LYMPH %/#, MIXED %/# (EO, BSO, MONO).
Status
Terminated
Initiated
2013-05-15
Posted
2013-10-18
Terminated
2016-08-15
Reason
The data printed on the thermal printer may omit a digit or decimal point in the results or sample identification number.
Root cause
Nonconforming Material/Component
Action
The firm, Sysmex, sent a "Product Notification-FIELD CORRECTION" letter dated May 2013 to all customers in the U.S. and Canada on May 15, 2013 by U.S. Mail. The letter describes the product, problem and actions to be taken. The customers will be contacted by a Sysmex Service Representative within the following 90 days to schedule an appointment to replace the Printed Circuit Board (PCB) on the internal printer. The repair will be performed during an on-site service visit. The customers were instructed to, until the repair can be made, use the thermal printer for reporting results; to compare the sample identification, values and decimal points on the printout with the data on the LCD screen for accuracy. If you have any questions concerning this information, please email the Sysmex Technical Assistance Center (TAC). Simply click on email TAC button on the home page of the CRC, complete the form that appears and then click the submit to send your question to TAC. Sending an email to TAC is recommended for non-urgent requests. For urgent request, call the Technical Assistance Center at 1-888-879-7639.
Quantity
193 units
Distribution
Worldwide distribution: US (Nationwide) including Puerto Rico and states of: AK, AL, AR, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, SD, TN, TX, UT, VA, VT, WA, WI, and WY; and country of Canada.
Lot, serial or model codes
Product Code: 023-1861-8;  Serial Numbers: B1677 - B1826, F1827 - F2161, F2178 - F2198, F2216 - F2396, F2411 - F2452, F2464 - F2472, F2495 - F2523, F2533, F2550 - F2560, F2592 - F2601, F2614 - F2635, F2640 - F2648, F2662 - F2679, F2710 - F2853, F2855 - F2871, F2878 - F2922, F2945, F2946, F2962 - F2970, F2998 - F3008, F3072 - F3080, F3083 - F3098
510(k) numbers
["K032677"]
PMA numbers
not on file
Product code
GKZ
Firm FEI
3009711478
Firm city
Lincolnshire
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register