Skip to the content

FDA Medical Device Recalls (openFDA)

Karl Storz Endoscopy: STORZ High Flow Insufflation Tubing (with filter), Catalog #20400161S/10 (outer box label), ENDOFLATOR 26-430520-1 also referenced on outer box), individually packaged in a sterile pouch (Catalog #20400161S), 10 pouches per cardboard box. Some of the pouches inside of cardboard box are incorrectly labeled as Catalog number 20400162S, High Flow Insufflation Tubing (with filter) For 26432020-1 THERMOFLATOR. The firm name on the label is Karl Storz Endoscopy-America, Inc., El Segundo, CA.

Recall number
Z-0098-2018
Recalling firm
Karl Storz Endoscopy
Product
STORZ High Flow Insufflation Tubing (with filter), Catalog #20400161S/10 (outer box label), ENDOFLATOR 26-430520-1 also referenced on outer box), individually packaged in a sterile pouch (Catalog #20400161S), 10 pouches per cardboard box. Some of the pouches inside of cardboard box are incorrectly labeled as Catalog number 20400162S, High Flow Insufflation Tubing (with filter) For 26432020-1 THERMOFLATOR. The firm name on the label is Karl Storz Endoscopy-America, Inc., El Segundo, CA.
Status
Terminated
Initiated
2017-10-10
Posted
not on file
Terminated
2018-06-22
Reason
The label outside the cardboard box is labeled correctly but the pouches inside of some of the boxes were mislabeled as a different product.
Root cause
Labeling mix-ups
Action
The recalling firm issued letters flagged as 'SAFETY ALERT" dated 10/5/2017 via Federal Express on 10/10/2017.
Quantity
496/10-pouch boxes
Distribution
Distribution was made nationwide. There was government distribution and no military distribution. Foreign distribution was made to Canada.
Lot, serial or model codes
Lot #129241
510(k) numbers
not on file
PMA numbers
not on file
Product code
NKC
Firm FEI
3007475226
Firm city
El Segundo
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register