FDA Medical Device Recalls (openFDA)
Small Bone Innovations, Inc.: Small Bone Innovations (SBi) rHead Radial Stem, rHead Recon Radial System, rHead Radial Stem Lateral Assembly, rHead Radial Stem 6mm Collar, rHead Recon Radial Stem 6mm Collar, rHead Radial Stem Lateral Assembly 6mm Collar, rHead Standard Extended Stem, rHead Radial Stem Plasma Coated, rHead Recon Radial Stem Plasma Coated, rHead Radial Stem 6mm Collar Plasma Coated, rHead Recon Radial Stem 6mm Collar Plasma Coated, uHead Ulnar Stem Standard, uHead Ulnar Stem 20mm Collar packaged in a sterile pouch inside a shrink-wrapped box. The recall was expanded on December 20, 2013 to include the following devices: rHead lateral assembly, radial head implant, sizes 2 through 4, Uni-elbow lateral assembly, radial implant assembly, sozes 2 through 4, ReMotion Left, large distal radial component, x-small, small, medium, and large, ReMotion Right, distal radial component, x-small, small, medium, and large.
- Recall number
- Z-0098-2014
- Recalling firm
- Small Bone Innovations, Inc.
- Product
- Small Bone Innovations (SBi) rHead Radial Stem, rHead Recon Radial System, rHead Radial Stem Lateral Assembly, rHead Radial Stem 6mm Collar, rHead Recon Radial Stem 6mm Collar, rHead Radial Stem Lateral Assembly 6mm Collar, rHead Standard Extended Stem, rHead Radial Stem Plasma Coated, rHead Recon Radial Stem Plasma Coated, rHead Radial Stem 6mm Collar Plasma Coated, rHead Recon Radial Stem 6mm Collar Plasma Coated, uHead Ulnar Stem Standard, uHead Ulnar Stem 20mm Collar packaged in a sterile pouch inside a shrink-wrapped box. The recall was expanded on December 20, 2013 to include the following devices: rHead lateral assembly, radial head implant, sizes 2 through 4, Uni-elbow lateral assembly, radial implant assembly, sozes 2 through 4, ReMotion Left, large distal radial component, x-small, small, medium, and large, ReMotion Right, distal radial component, x-small, small, medium, and large.
- Status
- Terminated
- Initiated
- 2013-09-06
- Posted
- 2013-10-29
- Terminated
- 2015-03-30
- Reason
- The recall was initiated due to an increased risk of implants breaching the sterile pouches inside the shrink-wrapped box.
- Root cause
- Packaging change control
- Action
- Email correspondence and a notification letter, dated September 6, 2013, were sent to users requesting the immediate return of affected devices. Another notification letter, dated December 20, 2013 was also sent for recall of the additional devices after the recall was expanded.
- Quantity
- 2036 devices
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of Canada and Mexico.
- Lot, serial or model codes
- all item numbers and lot numbers.
- 510(k) numbers
- ["K002644","K010786","K011819","K062898","K070236","K982288"]
- PMA numbers
- not on file
- Product code
- KWI
- Firm FEI
- 3003640913
- Firm city
- Morrisville
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28