Skip to the content

FDA Medical Device Recalls (openFDA)

Smith & Nephew, Inc.: smith&nephew TRIGEN INTERTAN 10S, REF 71675384, 10 MM X 18 CM NAIL, 130o (degrees), SET SCREW PREASSEMBLED IN NAIL

Recall number
Z-0097-2018
Recalling firm
Smith & Nephew, Inc.
Product
smith&nephew TRIGEN INTERTAN 10S, REF 71675384, 10 MM X 18 CM NAIL, 130o (degrees), SET SCREW PREASSEMBLED IN NAIL
Status
Open, Classified
Initiated
2017-10-24
Posted
not on file
Terminated
not on file
Reason
A single lot of INTERTAN 10S 10 MM X 18 CM 130 D was manufactured with the screw inserted upside down.
Root cause
Employee error
Action
Smith & Nephew sent an Urgent Medical Device Recall Notice dated October 24, 2017, to all affected customers. The firm requested return of the product. Please follow the instructions on the attached Response Form. For further questions, please call (978) 749-1440.
Quantity
17 units
Distribution
US Distribution - TN, IL, CT, MO, CA., and Internationally to Italy and Germany
Lot, serial or model codes
Lot number 16LM07111
510(k) numbers
["K040212"]
PMA numbers
not on file
Product code
JDS
Firm FEI
1020279
Firm city
Memphis
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register