Roche Diagnostics Operations, Inc.: cobas 6000 Series system c6000, Chemistry (Photometric, Discrete), for clinical Use. Modular e601: Immunoassay Analyzer Product Usage: The cobas p501/p701 instrument stores primary and secondary sample tubes. The tubes are stored in a controlled, refrigerated sample store. Automatic retrieval for add-on testing is possible at any time. Expired samples are disposed of automatically when their storage time has elapsed. The instrument is intended as accessory for diagnostic use.
- Recall number
- Z-0096-2017
- Recalling firm
- Roche Diagnostics Operations, Inc.
- Product
- cobas 6000 Series system c6000, Chemistry (Photometric, Discrete), for clinical Use. Modular e601: Immunoassay Analyzer Product Usage: The cobas p501/p701 instrument stores primary and secondary sample tubes. The tubes are stored in a controlled, refrigerated sample store. Automatic retrieval for add-on testing is possible at any time. Expired samples are disposed of automatically when their storage time has elapsed. The instrument is intended as accessory for diagnostic use.
- Status
- Terminated
- Initiated
- 2016-07-15
- Posted
- not on file
- Terminated
- 2017-06-05
- Reason
- Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.
- Root cause
- Device Design
- Action
- On 7/15/2016, URGENT MEDICAL DEVICE CORRECTION notifications were sent to the affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, an updated cleaning procedure, and instructions for responding to the formal recall notification. Please contact the Roche Support Network Customer Support Center at 1-800-428-2336 if you have questions about the information contained in this UMDC. On 10/5/2016, URGENT MEDICAL DEVICE CORRECTION UPDATE notifications were sent to the affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. The notification updated the 7/15/16 notification to include additional affected systems and to provide updated cleaning instructions for all systems via the enclosed Operators Manual addendum.
- Quantity
- 9014 in total
- Distribution
- US Nationwide Distribution
- Lot, serial or model codes
- Not applicable
- 510(k) numbers
- ["K060373"]
- PMA numbers
- not on file
- Product code
- JJE
- Firm FEI
- 1823260
- Firm city
- Indianapolis
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28