Skip to the content

FDA Medical Device Recalls (openFDA)

Codman & Shurtleff, Inc.: Surgical Pattie, 1/4" x 1/4" Non--Sterile (Each Unit of sales is 100 cards) of patties on each card) Product code: 24-5403

Recall number
Z-0095-2011
Recalling firm
Codman & Shurtleff, Inc.
Product
Surgical Pattie, 1/4" x 1/4" Non--Sterile (Each Unit of sales is 100 cards) of patties on each card) Product code: 24-5403
Status
Terminated
Initiated
2010-07-16
Posted
2010-10-20
Terminated
2012-07-09
Reason
Surgical Pattie x-ray stripe (radio-opaque marker) may become detached
Root cause
Process control
Action
Codman notified customers by letter dated 7/16/10 issued July 23 through July 26, 2010 via Fed'x overnight advising users that the x-ray marker on Codman Surgical Patties may become detached. Customers are instructed to return product from affected lots. An error in the original letter was discovered and a corrected letter dated July 28, 2010 issued. If there are any questions, please contact Codman at (800) 225-0460.
Quantity
229 boxes (100 cards/10 patties each card)
Distribution
Nationwide Foreign: Canada, Singapore, Belgium, Mexico, Madrid, Dubai, Tokyo, Brazil, New Zealand, Egypt, South Africa, Singapore,
Lot, serial or model codes
09A0061 09B0310 09C0708  09C0709  09C0709  09E1139  09F1516 09F1517  09G1830  09G1830  09H2170  09J2622  09H2170  09J2622  09L3080 10B0196 10B0317  10C0442  10B0318  10C0442 Lot numbers:  8L2688
510(k) numbers
["K880402"]
PMA numbers
not on file
Product code
HBA
Firm FEI
3002947440
Firm city
Raynham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register