Skip to the content

FDA Medical Device Recalls (openFDA)

Mckesson Information Solutions LLC: McKesson Cardiology Hemo. McKesson Cardiology" Hemo is intended for complete physiological/hemodynamic monitoring, clinical data acquisition, medical image and data processing, and analytical assessment.

Recall number
Z-0093-2015
Recalling firm
Mckesson Information Solutions LLC
Product
McKesson Cardiology Hemo. McKesson Cardiology" Hemo is intended for complete physiological/hemodynamic monitoring, clinical data acquisition, medical image and data processing, and analytical assessment.
Status
Terminated
Initiated
2014-06-25
Posted
2014-10-20
Terminated
2017-10-26
Reason
McKesson Cardiology Hemo front-end parameter boxes must undergo annual maintenance testing to establish the proper functioning of the parameter boxes, and to identify in a timely manner whether a parameter box is out of calibration. McKesson has identified that its communication to customers regarding the required maintenance tests was insufficient. While each unit has an attached sticker i
Root cause
Labeling design
Action
Consignees were notified via Field Safety Notice on 06/25/2014.
Quantity
625 devices
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Australia, Botswana, Canada, Germany, New Zealand, Netherlands, Pakistan, Portugal, South Africa, Turkey and The United Kingdom.
Lot, serial or model codes
McKesson Cardiology TM Hemo versions: 11.0HF5, 11.1, 11.1SP5, 11.1SP6, 11.1SP6HF1, 11.1SP6HF3, 12SP1, 12.1, 12.1HF1, 12.1HF2, 12.2, 12.2HF1, 12.2HF2, 12.2HF3, 12.2HF4.
510(k) numbers
["K131497"]
PMA numbers
not on file
Product code
MWI
Firm FEI
3002684279
Firm city
Alpharetta
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register