B. Braun Medical, Inc.: Extension Set, EXT SET W/CARESITE, 6IN., Product Code 470169, 50 units per carton. A6 Access Device is a valve intended for the aspiration, injection or gravity/pump flow of IV fluids and blood upon insertion of a male luer connector.

Recall number
Z-0091-2023
Recalling firm
B. Braun Medical, Inc.
Product
Extension Set, EXT SET W/CARESITE, 6IN., Product Code 470169, 50 units per carton. A6 Access Device is a valve intended for the aspiration, injection or gravity/pump flow of IV fluids and blood upon insertion of a male luer connector.
Status
Open, Classified
Initiated
2022-09-15
Posted
2022-10-14
Terminated
not on file
Reason
The extension set has a high probability of failure which could lead to possible leakage or detachment at the device's bonded joint during use. The issue may result in delay of therapy, increased risk of bloodstream infections (BSI), and may expose both patients and healthcare workers to hazardous medications. Depending on the nature of the leak and type of infusion, there is also the potential for life-threatening situations to occur including air embolization or hypovolemia.
Root cause
Process design
Action
On September 15, 2022, the firm notified affected customers via Urgent Medical Device Recall letters. Customers were instructed to cease use and quarantine the product subject to recall. Once customers have returned the completed acknowledgement form to B. Braun Medical Inc., a Customer Support representative will contact the customer with instructions on returning impacted cases. The firm will provide credit and/or replacement of the product based on the customer's needs. For questions, please contact BBMI's Postmarket Surveillance Department at 1-833-425-1464.
Quantity
354,200 US; 1,800 OUS
Distribution
Distribution US nationwide, Canada and Saudi Arabia.
Lot, serial or model codes
UDI-DI (GTIN): 04046964540943 Lots/Expiration Dates 61705083 31-Oct-22 61705124 31-Oct-22 61706055 31-Oct-22 61714103 31-Dec-22 61719696 31-Jan-23 61732456 31-Mar-23 61736580 31-May-23 61739427 31-May-23 61744671 31-Jul-23 61747367 31-Aug-23 61750206 31-Aug-23 61752658 30-Sep-23 61761422 30-Nov-23 61776447 30-Apr-24 61780880 30-Apr-24 61802320 30-Sep-24 61806067 31-Oct-24 61807459 31-Oct-24 61822282 28-Feb-25
510(k) numbers
["K083723","K942988","K954123"]
PMA numbers
not on file
Product code
FPA
Firm FEI
2523676
Firm city
Allentown
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register