FDA Medical Device Recalls (openFDA)
Ortho-Clinical Diagnostics: VITROS Chemistry Products TRIG Slides, REF/Product Codes 133 6544 (300 slides per sales unit) and 832 9930 (90 slides per sales unit), Unique Device Identifier Nos. 10758750008889 and 10758750004768, Rx ONLY, IVD --- For in vitro diagnostic use only. VITROS Chemistry Products TRIG Slides quantitatively measure triglyceride (TRIG) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System. Triglyceride measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders
- Recall number
- Z-0091-2017
- Recalling firm
- Ortho-Clinical Diagnostics
- Product
- VITROS Chemistry Products TRIG Slides, REF/Product Codes 133 6544 (300 slides per sales unit) and 832 9930 (90 slides per sales unit), Unique Device Identifier Nos. 10758750008889 and 10758750004768, Rx ONLY, IVD --- For in vitro diagnostic use only. VITROS Chemistry Products TRIG Slides quantitatively measure triglyceride (TRIG) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System. Triglyceride measurements are used in the diagnosis and treatment of patients with diabetes mellitus, nephrosis, liver obstruction, other diseases involving lipid metabolism, or various endocrine disorders
- Status
- Terminated
- Initiated
- 2016-08-11
- Posted
- not on file
- Terminated
- 2018-06-11
- Reason
- Ortho Clinical Diagnostics confirmed that N-Acetylcysteine (NAC), a drug used for acetaminophen overdose, mucolytic therapy and is used as nutritional supplement, can interfere with VITROS GLU, LAC, TRIG, and URIC Slides assays.
- Root cause
- Under Investigation by firm
- Action
- Ortho-Clinical Diagnostics sent an URGENT PRODUCT CORRECTION Letter (Ref. CL2016-152) dated August 11, 2016 via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/or US Postal Service Priority Mail (for PO Boxes only) to all customers who could order and use affected product to inform them that new interferant were identified for GLU, LAC, TRIG and URIC Slides which may cause biased results. Foreign affiliates were informed by e-mail on August 11, 2016, of the issue and instructed to notify their consignees of the issue. For questions, please contact our Ortho CareTM Technical Solutions Center at 1-800-421-3311. For questions regarding this recall call 908-218-8776.
- Quantity
- US: 60,372 sales units; Foreign: 145,611 sales units
- Distribution
- Worldwide Distribution - US (Nationwide) Puerto Rico and Bermuda), Internationally to Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Poland, Singapore, Spain, and Venezuela.
- Lot, serial or model codes
- Current/In-date GENs 16, 17, 19, 20, 22, 23, 24, 25, 27 through 32
- 510(k) numbers
- ["K130332"]
- PMA numbers
- not on file
- Product code
- CDT
- Firm FEI
- 1000136573
- Firm city
- Rochester
- Firm state
- NY
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28