Carestream Health, Inc.: DRX Revolution Mobile X-Ray System, Model DRXR-1, Verona Street Rochester, NY 14608 Catalog Numbers 1019397, 1023415, 1023423 --- Made in USA by Carestream Health, Inc. 150 The device is designed to perform radiographic x-ray examinations on all pediatric and adult patients, in all patient treatment areas.
- Recall number
- Z-0090-2014
- Recalling firm
- Carestream Health, Inc.
- Product
- DRX Revolution Mobile X-Ray System, Model DRXR-1, Verona Street Rochester, NY 14608 Catalog Numbers 1019397, 1023415, 1023423 --- Made in USA by Carestream Health, Inc. 150 The device is designed to perform radiographic x-ray examinations on all pediatric and adult patients, in all patient treatment areas.
- Status
- Terminated
- Initiated
- 2013-07-26
- Posted
- 2013-10-31
- Terminated
- 2016-05-25
- Reason
- During an audit, Carestream discovered a DRX-1 System that captured an image which displayed an artifact. This issue was discovered in manufacturing. It occurs when the DRX product console software is updated to v5.6b software and the detector is not calibrated before use.
- Root cause
- Software design
- Action
- Carestream sent a Urgent-Medical Device Recall letter dated July 26, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Carestream Health Inc. recently discovered an image artifact issue due to a detector not being recalibrated with the 5.6B software upgrade. As our installation instructions did not require the user to recalibrate during an upgrade, there is a possibility your detector could do the same. When this occurs an image artifact could be evident on images captured in the Extended Exposure mode. These artifacts could potentially contribute to misdiagnosis or the necessity of patient re-exposure. We request that you take a moment to verify that your detector(s) are calibrated. At our next visit, Carestream will also verify that your detector was re-calibrated and, if not, re-calibrate the detector for you. Please note this only affects systems upgraded in the field. For questions or concerns, please contact the Carestream Customer Care Center at 1-800-328-2910, 8 am through 8 pm EST in the US.
- Quantity
- 23 units installed total (11 domestically & 12 internationally)
- Distribution
- Worldwide Distribution - US (nationwide) including the states of New York, Massachusetts, Pennsylvania, Illinois, Texas, California and Louisiana., and the countries of Australia, Canada, Finland, Italy, Switzerland and United Kingdom.
- Lot, serial or model codes
- Serial numbers: U.S. - 134,152,173,174,200,205,222,312,349,353,412, Worldwide Serial numbers: 469, 445, 306, 410, 453, 204, 483, 300, 171, 218
- 510(k) numbers
- ["K120062"]
- PMA numbers
- not on file
- Product code
- IZL
- Firm FEI
- 1315356
- Firm city
- Rochester
- Firm state
- NY
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28