Siemens Healthcare Diagnostics, Inc.: Dimension Vista High Density Lipoprotein Cholesterol (HDLC) Flex reagent cartridge, IVD Siemens Material Number (SMN): 10464340 Catalog Number: K3048A
- Recall number
- Z-0089-2023
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Product
- Dimension Vista High Density Lipoprotein Cholesterol (HDLC) Flex reagent cartridge, IVD Siemens Material Number (SMN): 10464340 Catalog Number: K3048A
- Status
- Open, Classified
- Initiated
- 2022-08-24
- Posted
- 2022-10-14
- Terminated
- not on file
- Reason
- Dimension Vista High Density Lipoprotein Cholesterol (HDLC) Potential for Erroneous Result. If an HDLC QC or patient result is obtained without an Abnormal Reaction [E145]: flag, a falsely depressed or elevated result ranging from -90% to 133% may be observed.
- Root cause
- Under Investigation by firm
- Action
- Siemens issued an Urgent Medical Device Correction (UMDC) mailed to US affected customers by FedEx beginning on 08/24/2022. An Urgent Field Safety Notice (UFSN) was issued to all OUS Siemens reporting sites with affected customers, for distribution and regional reporting to their local competent authorities beginning on 08/24/2022. Letter states: Discontinue use of and discard the lot listed in Table 1. Review your inventory of this lot to determine your laboratory s replacement needs and to provide information to Siemens for reporting to the Authorities. Please review this letter with your Medical Director. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days to Siemens Healthineers for reporting to the Authorities. If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers Technical Support Representative. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers Technical Support Representative.
- Quantity
- 1592 units
- Distribution
- Nationwide Foreign: Canada
- Lot, serial or model codes
- UDI-DI : 0084276802043821250BA2022090710464340840 Lot Number: 21250BA Exp. Date: 2022-09-07
- 510(k) numbers
- ["K081300"]
- PMA numbers
- not on file
- Product code
- JHM
- Firm FEI
- 2517506
- Firm city
- Newark
- Firm state
- DE
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28