Codman & Shurtleff, Inc.: Codman EDS 3 Drainage System with Ventricular Catheter (External Drainage System) Catalog Number: 82-1730 Indicated for draining cerebrospinal fluid (CFS) and other fluids of similar physical characteristics as a means of reducing intracranial pressure and CSF volume when the insertion of a permanent, internal shunt is not indicated

Recall number
Z-0089-2015
Recalling firm
Codman & Shurtleff, Inc.
Product
Codman EDS 3 Drainage System with Ventricular Catheter (External Drainage System) Catalog Number: 82-1730 Indicated for draining cerebrospinal fluid (CFS) and other fluids of similar physical characteristics as a means of reducing intracranial pressure and CSF volume when the insertion of a permanent, internal shunt is not indicated
Status
Terminated
Initiated
2014-09-15
Posted
2014-10-20
Terminated
2017-08-11
Reason
tubing within the system that drains CSF may leak or disconnect from the joints and may result in over- or under-drainage of CSF from the ventricular system or introduction of air into the ventricular system (pneumocephalus)
Root cause
Device Design
Action
Codman Neuro sent an Urgent Notice dated September 12, 2014, to all affected customers via overnight mailing and OUS customers notified via email the week of September 15, 2014. Accounts were requested to isolate all inventory of the affected product and return to Stericycle at the following address: Stericycle 2670 Executive Drive, Suite A Attn: Event #8894 Indianapolis, IN 46241 For patients currently being managed with the EDS 3 System, the system should be replaced immediately. In the case where no substitute drainage system is immediately available, the EDS 3 System may continue to be used until an alternative product can be obtained. If you have any questions or concerns regarding this notification, please contact your local Codman Neuro Representative or Scientific and Medical Affairs at SciMedAffairs@its.jnj.com or (866) 685-7325. For questions regarding this recall call 508-880-8000.
Quantity
66,120 units
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to: United Arab Emirates Argentina Austria Australia Aruba Belgium Bahrain Bolivia Brazil Bahamas Canada Switzerland Chile People's Republic of China Columbia Costa Rica Cyprus Czech Republic Germany Algeria Estonia Egypt Spain Finland France Gabon Great Britain French Guiana Greece Guatemala Hong Kong Croatia Hungary Ireland Israel India Iran Italy Jordan Kenya Kuwait Kazakhstan Lebanon Lithuania Luxembourg Latvia Mauritius Mexico Malaysia Mozambique New Caledonia Nigeria Netherlands Norway New Zealand Oman Panama Peru Poland Palestine Portugal Reunion Romania Serbia Russian Federation Rwanda Saudi Arabia Sweden Singapore Slovenia Slovakia Tunisia Turkey Trinadad And Tobago Taiwan Ukraine Uruguay Venezuela Mayotte South Africa
Lot, serial or model codes
All lots with expiration on or before August 2017 (2017-08)
510(k) numbers
["K061568"]
PMA numbers
not on file
Product code
JXG
Firm FEI
3002947440
Firm city
Raynham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register