FDA Medical Device Recalls (openFDA)
Medrad Inc: Medrad Overhead Counterpoise System II. Support Arm. Mount Injector Head. Catalog numbers 3010544, 3010545, 3010546, 3010547, 3010548, 3010549, 3010550, 3010551, 3010552, and 3010553.
- Recall number
- Z-0089-06
- Recalling firm
- Medrad Inc
- Product
- Medrad Overhead Counterpoise System II. Support Arm. Mount Injector Head. Catalog numbers 3010544, 3010545, 3010546, 3010547, 3010548, 3010549, 3010550, 3010551, 3010552, and 3010553.
- Status
- Terminated
- Initiated
- 2005-10-07
- Posted
- 2005-10-25
- Terminated
- 2007-11-29
- Reason
- overhead system can fall due to the failure of a retaining ring.
- Root cause
- Other
- Action
- The recalling firm issued an Urgent Medical Device Field Correction notice dated 10/7/05 to all customers. The notice informs customers of the problem and cautions anyone to be under the system when in use. The notice also informs the customer that a firm representative will schedule a time to perform the upgrade. On 10/13/05, the recalling firm also issued a notification to scanner manufacturers alerting them of the problem.
- Quantity
- 2080 units
- Distribution
- The product was shipped to medical facilities nationwide. The product was also shipped to Austria, Belgium, Chile, Czech, Denmark, Spain, Finland, France, Great Britain, Ireland, Israel, Italy, Netherlands, Norway, Sweden, Austrailia, Kuwait, Saudia Arabia, Canada, China, Hong Kong, India, Korea, Singapore, Taiwan, Brazil, Columbia, Cuba, Mexico, and Japan.
- Lot, serial or model codes
- All serial numbers
- 510(k) numbers
- ["K944983"]
- PMA numbers
- not on file
- Product code
- JAK
- Firm FEI
- 2520313
- Firm city
- Indianola
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28