Skip to the content

FDA Medical Device Recalls (openFDA)

Medrad Inc: Medrad Overhead Counterpoise System II. Support Arm. Mount Injector Head. Catalog numbers 3010544, 3010545, 3010546, 3010547, 3010548, 3010549, 3010550, 3010551, 3010552, and 3010553.

Recall number
Z-0089-06
Recalling firm
Medrad Inc
Product
Medrad Overhead Counterpoise System II. Support Arm. Mount Injector Head. Catalog numbers 3010544, 3010545, 3010546, 3010547, 3010548, 3010549, 3010550, 3010551, 3010552, and 3010553.
Status
Terminated
Initiated
2005-10-07
Posted
2005-10-25
Terminated
2007-11-29
Reason
overhead system can fall due to the failure of a retaining ring.
Root cause
Other
Action
The recalling firm issued an Urgent Medical Device Field Correction notice dated 10/7/05 to all customers. The notice informs customers of the problem and cautions anyone to be under the system when in use. The notice also informs the customer that a firm representative will schedule a time to perform the upgrade. On 10/13/05, the recalling firm also issued a notification to scanner manufacturers alerting them of the problem.
Quantity
2080 units
Distribution
The product was shipped to medical facilities nationwide. The product was also shipped to Austria, Belgium, Chile, Czech, Denmark, Spain, Finland, France, Great Britain, Ireland, Israel, Italy, Netherlands, Norway, Sweden, Austrailia, Kuwait, Saudia Arabia, Canada, China, Hong Kong, India, Korea, Singapore, Taiwan, Brazil, Columbia, Cuba, Mexico, and Japan.
Lot, serial or model codes
All serial numbers
510(k) numbers
["K944983"]
PMA numbers
not on file
Product code
JAK
Firm FEI
2520313
Firm city
Indianola
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register