FDA Medical Device Recalls (openFDA)
Draeger Medical, Inc.: Draeger Medical Apollo Anesthesia Machine, catalog number 8606500
- Recall number
- Z-0087-2009
- Recalling firm
- Draeger Medical, Inc.
- Product
- Draeger Medical Apollo Anesthesia Machine, catalog number 8606500
- Status
- Terminated
- Initiated
- 2008-08-22
- Posted
- 2008-10-08
- Terminated
- 2009-09-01
- Reason
- Sporadic errors in various device functions, including low readings of the single gas flows for oxygen, nitrous oxide, or air delivery, or mechanical ventilation failure. Alarms and error messages function as designed.
- Root cause
- Nonconforming Material/Component
- Action
- Draeger Medical sent a Medical Device Recall letter to their customers via certified mail on 8/22/08 informing them of the problem. Until the exchange of affected printed circuit boards has been implemented, the firm is recommending that users closely monitor all device displays and error messages and if applicable act in accordance with the advisory statements and remedial information provided in the Instructions for Use as well as making provisions for alternative patient care measures.
- Quantity
- 214 units
- Distribution
- Nationwide Distribution.
- Lot, serial or model codes
- Catalog number 8606500. Serial numbers ARZC-0142, ARZC-0143, ARZC-0144, ARZC-0145, ARZC-0146, ARZC-0147, ARZC-0148, ARZC-0149, ARZC-0150, ARZC-0151, ARZC-0152, ARZC-0153, ARZC-0172, ARZC-0173, ARZC-0174, ARZC-0175, ARZC-0176, ARZC-0177, ARZC-0192, ARZC-0193,ARZC-0196, ARZC-0197, ARZC-0198, ARZC-0199, ARZC-0200, ARZD-0001, ARZD-0002, ARZD-0003, ARZD-0010, ARZD-0011, ARZD-0012, ARZD-0013, ARZD-0014, ARZD-0015, ARZD-0016, ARZD-0017, ARZD-0018, ARZD-0019, ARZD-0020, ARZD-0021, ARZD-0022, ARZD-0023, ARZD-0024, ARZD-0025, ARZD-0026, ARZD-0027, ARZD-0028, ARZD-0029, ARZD-0030, ARZD-0031, ARZD-0032, ARZD-0033, ARZD-0034, ARZD-0035, ARZD-0036, ARZD-0037, ARZD-0038, ARZD-0039, ARZD-0040, ARZD-0041, ARZD-0042, ARZD-0043, ARZD-0044, ARZD-0058, ARZD-0059, ARZD-0060, ARZD-0061, ARZD-0062, ARZD-0063, ARZD-0064, ARZD-0065, ARZD-0066, ARZD-0067, ARZD-0068, ARZD-0069, ARZD-0070, ARZD-0071, ARZD-0072, ARZD-0073, ARZD-0074, ARZD-0075, ARZD-0076, ARZD-0077, ARZD-0078, ARZD-0079, ARZD-0080, ARZD-0081, ARZD-0082, ARZD-0083, ARZD-0084, ARZD-0085, ARZD-0086, ARZD-0087, ARZD-0117, ARZD-0118, ARZD-0119, ARZD-0120, ARZD-0121, ARZD-0122, ARZD-0125, ARZD-0126, ARZD-0127, ARZD-0128, ARZD-0129, ARZD-0130, ARZD-0131, ARZD-0132, ARZD-0133, ARZD-0134, ARZD-0135, ARZD-0136, ARZD-0183, ARZD-0184, ARZD-0185, ARZD-0186, ARZD-0187, ARZD-0188, ARZD-0189, ARZD-0190, ARZD-0191, ARZD-0192, ARZD-0193, ARZD-0194, ARZD-0195, ARZD-0196, ARZD-0197, ARZD-0198, ARZD-0199, ARZD-0200, ARZD-0201, ARZD-0202, ARZD-0203, ARZD-0255, ARZD-0256, ARZD-0257, ARZD-0258, ARZD-0259, ARZD-0260, ARZD-0261, ARZD-0262, ARZD-0263, ARZD-0264, ARZD-0265, ARZD-0266, ARZD-0267, ARZD-0268, ARZD-0269, ARZD-0270, ARZD-0271, ARZE-0001, ARZE-0002, ARZE-0003, ARZE-0004, ARZE-0005, ARZE-0008, ARZE-0009, ARZE-0010, ARZE-0023, ARZE-0024, ARZE-0025, ARZE-0026, ARZE-0027, ARZE-0028, ARZE-0029, ARZE-0030, ARZE-0059, ARZE-0060, ARZE-0061, ARZE-0062, ARZE-0063, ARZE-0064, ARZE-0065, ARZE-0115, ARZE-0116, ARZE-0173, ARZE-0174, ARZE-0175, ARZE-0176, ARZE-0189, ARZE-0190, ARZE-0191, ARZE-0192, ARZE-0193, ARZE-0194, ARZE-0195, ARZE-0196, ARZE-0197, ARZE-0198, ARZE-0199, ARZE-0200, ARZE-0201, ARZE-0202, ARZE-0203, ARZE-0204, ARZE-0214, ARZE-0215, ARZE-0216, ARZF-0011, ARZF-0012, ARZF-0013, ARZF-0014, ARZF-0015, ARZF-0016, ARZF-0017, ARZF-0018, ARZF-0019, ARZF-0020, ARZF-0021, ARZF-0022, ARZF-0023, ARZF-0024, ARZF-0025, ARZF-0026, ARZF-0027, ARZF-0028, ARZF-0029, ARZF-0030, ARZF-0031, ARZF-0084, ARZF-0087, ARZF-0088, ARZF-0089, ARZF-0090, ARZF-0091, ARZF-0092, ARZF-0093, ARZF-0094, ARZF-0095, ARZF-0096, ARZF-0097, ARZF-0098, ARZF-0099, ARZF-0100, ARZF-0101, ARZF-0102, ARZF-0103, ARZF-0104, ARZF-0105, ARZF-0106, ARZF-0107, ARZF-0108, ARZF-0109, ARZF-0110, and ARZF-0111.
- 510(k) numbers
- ["K042607"]
- PMA numbers
- not on file
- Product code
- BSZ
- Firm FEI
- 2517967
- Firm city
- Telford
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28