Skip to the content

FDA Medical Device Recalls (openFDA)

Roche Diagnostics Operations, Inc.: DISCOVERY ULTRA Advanced Staining System (RUO-Research Use Only) Material Number: 05987750001

Recall number
Z-0086-2021
Recalling firm
Roche Diagnostics Operations, Inc.
Product
DISCOVERY ULTRA Advanced Staining System (RUO-Research Use Only) Material Number: 05987750001
Status
Open, Classified
Initiated
2020-08-31
Posted
not on file
Terminated
not on file
Reason
Potential for Fluid leaks within the instruments, creating an electrical short inside the power socket, which generates heat, smoke, and eventually burning and melting of the electrical cord connector and socket
Root cause
Device Design
Action
Roche issued Urgent Medical Device Correction letter on 8/31/20 via UPS Ground Receipt stating: reason for recall, health risk and action to take: actions below. " Inform any operators that use the BenchMark ULTRA and DISCOVERY ULTRA instruments of the potential hazards associated with this issue and provide a copy of this notification as appropriate. " Inspect the instrument daily for evidence of leaks (i.e., liquid outside of the waste reservoir, liquid on the floor, liquid on the skin/outer panel of the instrument). " If you experience a fluid leak, stop using the system and turn off the power to the instrument. Immediately contact the Roche Support Network Customer Support Center at 1-800-227-2155. " Complete the attached fax back form (TP-01057) and fax or email it according to the instructions on the form. " If your facility has distributed the affected product to another site, please ensure this UMDC is provided to that site. " File this UMDC for future reference Local Field Engineering Specialists (FESs) will perform specific service actions, including the completion of the Waste Tub and Filter Upgrade (if it has not already been implemented), inspection of specific instrument parts for signs of cracks and fluid leaks, and installation of a power connector shield, which will prevent fluid entry into the power cord connection and socket. Contact the Roche Support Network Customer Support Center 24 hours a day, seven days a week at 1-800-227-2155 if you have questions about the information contained in this UMDC.
Quantity
257 (US); 208 (ex-US)
Distribution
Nationwide Foreign: Canada, Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Ireland, Italy, Latvia, Lithuania, Luxembourg, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, UK, Algeria, Argentina, Aruba, Australia, Bahrain, Bosnia-Herz., Brazil, Chile, China, Colombia, Costa Rica, Cuba, Dominican Republic, Ecuador, Egypt, El Salvador, Hong Kong, India, Israel, Japan, Jordan, Kazakhstan, Kuwait, Malaysia, Martinque, Mexico, Monaco, Morocco, Namibia, New Caledonia, New Zealand, Oman, Pakistan, Panama, Peru, Philippines, Russia, Saudi Arabia, Singapore, S. Africa, S. Korea, Taiwan, Thailand, Ukraine, UAE, Vietnam
Lot, serial or model codes
All units
510(k) numbers
not on file
PMA numbers
not on file
Product code
KPA
Firm FEI
1823260
Firm city
Indianapolis
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register