Ortho Development Corporation: Ortho Development Corp., Pivot BiPolar Head, 28mm, Hip Prosthesis, Femoral Head.

Recall number
Z-0086-2010
Recalling firm
Ortho Development Corporation
Product
Ortho Development Corp., Pivot BiPolar Head, 28mm, Hip Prosthesis, Femoral Head.
Status
Terminated
Initiated
2009-04-29
Posted
2009-10-28
Terminated
2010-05-25
Reason
A Pivot BiPolar Head, the femoral head in a hip prosthesis, might not move freely after assembly due a stacking up of manufacturing tolerances.
Root cause
Device Design
Action
Ortho sent a Field notification Memo dated 4/2/2009 to agents and distributors, giving them pre-implant examination guidelines. Decision to recall non-implanted units was made on 4/29/2009 and this was communicated by telephone to distributors.
Quantity
23
Distribution
AZ, CA, FL, MA, MI. No military, government or foreign distribution.
Lot, serial or model codes
Part 133-2846, Lots 3370, 3534, 3620, 3731 Part 133-2847, Lots 3492, 3535 Part 133-2848, Lot 3509 Part 133-2851, Lot 3433 Part 133-2852, Lot 3661
510(k) numbers
["K050966"]
PMA numbers
not on file
Product code
KWY
Firm FEI
3017298928
Firm city
Draper
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register