Skip to the content

FDA Medical Device Recalls (openFDA)

ConMed Corporation: CONCEPT HEATWAVE Electrode 130 mm x 2.3 mm diameter, REF/Catalog Number CS-023, STERILE, Rx Only Product Usage: The HeatWave Electrode is intended to be used for electro-coagulation of soft tissue in shoulder, ankle, wrist, elbow and knee arthroscopic procedures.

Recall number
Z-0085-2017
Recalling firm
ConMed Corporation
Product
CONCEPT HEATWAVE Electrode 130 mm x 2.3 mm diameter, REF/Catalog Number CS-023, STERILE, Rx Only Product Usage: The HeatWave Electrode is intended to be used for electro-coagulation of soft tissue in shoulder, ankle, wrist, elbow and knee arthroscopic procedures.
Status
Terminated
Initiated
2016-08-29
Posted
not on file
Terminated
2018-06-07
Reason
CONMED received complaints of damage to the CS-023 electrodes shaft insulation which could result in burn marks on the shaft. If the shaft insulation is damaged in any way, it could increase the risk of patient burn.
Root cause
Process change control
Action
Conmed sent an URGENT DEVICE RECALL letter dated August 31, 2016 and Business Reply Form were sent to the domestic consignees on August 29, 2016 via USPS First Class Mail. The letter identified the affected product, problem and actions to be taken. For questions or requests contact the Field Action Support Team at 1-800-535-8536, fax to 315-624-3225, or email largheatwv@conmed.com.
Quantity
Domestic: 1740; Foreign: 4770
Distribution
Worldwide Distribution - US Nationwide in the states of: CA, HI, IL, IN, KS, LA, MA, NC, NM, NY, OH, OK, PA, TN, TX, UT, WA and WI; plus Foreign distribution to Australia, Canada, France, Germany, Italy, Spain, Czech Republic, Lithuania, Netherlands, Poland, Korea,Morocco, Malaysia, Mexico, Taiwan, and South Africa.
Lot, serial or model codes
All lots codes manufactured from 09/01/2011 through 11/17/2015: 267221, 301001, 309889, 318652, 319649, 325019, 354759, 334627, 336041, 341704, 343927, 349952, 384294, 354768, 365239, 369773, 374480, 378521, 406137, 385695, 388538, 392382, 394178, 396542, 436808, 413702, 456341, 423877, 427704, 433503, 478513, 451923, 488101, 464062, 468012, 471297, 511327, 482366, 522335, 494840, 499981, 505321, 544240, 515060, 551152, 527734, 531802, 536700, 581072, 550988, 590037, 557461, 570449, 574167, 611804, 588892, 621304, 597982, 598682, 605515, 643396, 619686, 656909, 629202, 634911, 643395, 676005, 654991, 692617, 667118, 671471, 674410, 683144, 695062, 695063, 297982.
510(k) numbers
["K991830"]
PMA numbers
not on file
Product code
JOS
Firm FEI
3001451571
Firm city
Utica
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register