FDA Medical Device Recalls (openFDA)
Go Medical Industries Pty., Ltd.: O'Neil Sterile Field Intermittent Urinary Catheter Kit labeled in mart:"***ORIGINAL CLOSED SYSTEM BY DR O'NEIL O'NEIL STERILE FIELD INTERMITTENT URINARY CATHETER KIT WITH IODINE SWABSTICKS (3)*** REF: ONSM 16KPES SIZE:16Fr Product No: AAM 8014KPES***SINGLE USE ONLY***Original design by: Dr. George O'Neil 200 Churchill Ave, Subiaco, WA 6008, Australia Tel.: +61 (0)8 9388 1700 Fax: +61 (0)8 9382 4952 European Representative: Advena Ltd Hereford, HR4 9DQ, UK DISTRIBUTED IN THE USA BY: American Australian Medical, Inc. 1-888-880-8871***" Case of Catheter Kits labeled in part:"***The O'Neil Sterile Field Urinary Catheter Kit REF: ONSM 16KPES Product No: AAM 8016KPES***CAUTION: Federal law restrict sale of this device to or on the order of a physician. STERILE Single use only***Designed by: GO MEDICAL INDUSTRIES PTY LTD 200 Churchill Avenue, Subiaco, WA 6008, Australia Tel.: +61 (0)8 9388 1700 Fax: +61 (0)8 9382 4952 email: gomedic@linet.net.au Website: www.gomedical.com.au***Distributed in the U.S.A. by: AMERICAN AUSTRALIAN MEDICAL INC. BUFFALO, NY 14209 USA TEL: 1-888-880-8871***Country of Origin: ROMANIA***QUANTITY 50***" This kit includes a Urinary Catheter and accessories including gloves, tissue, underpads, and Triad Povidone Iodine Swabsticks (3). These accessories are included in the kit as optional and are not required for urinary catheterization.
- Recall number
- Z-0085-2012
- Recalling firm
- Go Medical Industries Pty., Ltd.
- Product
- O'Neil Sterile Field Intermittent Urinary Catheter Kit labeled in mart:"***ORIGINAL CLOSED SYSTEM BY DR O'NEIL O'NEIL STERILE FIELD INTERMITTENT URINARY CATHETER KIT WITH IODINE SWABSTICKS (3)*** REF: ONSM 16KPES SIZE:16Fr Product No: AAM 8014KPES***SINGLE USE ONLY***Original design by: Dr. George O'Neil 200 Churchill Ave, Subiaco, WA 6008, Australia Tel.: +61 (0)8 9388 1700 Fax: +61 (0)8 9382 4952 European Representative: Advena Ltd Hereford, HR4 9DQ, UK DISTRIBUTED IN THE USA BY: American Australian Medical, Inc. 1-888-880-8871***" Case of Catheter Kits labeled in part:"***The O'Neil Sterile Field Urinary Catheter Kit REF: ONSM 16KPES Product No: AAM 8016KPES***CAUTION: Federal law restrict sale of this device to or on the order of a physician. STERILE Single use only***Designed by: GO MEDICAL INDUSTRIES PTY LTD 200 Churchill Avenue, Subiaco, WA 6008, Australia Tel.: +61 (0)8 9388 1700 Fax: +61 (0)8 9382 4952 email: gomedic@linet.net.au Website: www.gomedical.com.au***Distributed in the U.S.A. by: AMERICAN AUSTRALIAN MEDICAL INC. BUFFALO, NY 14209 USA TEL: 1-888-880-8871***Country of Origin: ROMANIA***QUANTITY 50***" This kit includes a Urinary Catheter and accessories including gloves, tissue, underpads, and Triad Povidone Iodine Swabsticks (3). These accessories are included in the kit as optional and are not required for urinary catheterization.
- Status
- Terminated
- Initiated
- 2011-09-03
- Posted
- 2011-10-21
- Terminated
- 2012-01-30
- Reason
- O'Neil Sterile Field Intermittent Urinary Catheter Kits With Povidone Iodine Swabsticks are being recalled due to a recall on the Triad Povidone Iodine Swabsticks contained within the kit. H&P Industries are recalling the Triad Povidone Iodine Swabsticks because it was determined that the swabsticks were manufactured without having a system for microbial testing in place at the time of release.
- Root cause
- Nonconforming Material/Component
- Action
- The firm, Go Medical Industries Pty Ltd, sent an e-mail on September 3, 2011 and follow-up with an "URGENT: MEDICAL DEVICE RECALL" letter dated September 07, 2011 to its customers. The letter described the product, problem and action to be taken. The customers were instructed to quarantine and not sell the product in the market place; immediately perform a sub-recall and notify their customers and complete and return the RETURN RESPONSE FORM via fax at 716-883-8099 or mail to: American Australian Medical Inc, 1083 Delaware Avenue, Buffalo, NY 14209. For further details you can contact QA Manager at Go Medical Industries at +61 89 381 3226 or pmoretta@gomedical.com.au.
- Quantity
- 14,700
- Distribution
- Nationwide distribution.
- Lot, serial or model codes
- Lot #'s: 9080 9081 9361 9362 9865 9866 9867
- 510(k) numbers
- ["K002803"]
- PMA numbers
- not on file
- Product code
- FCM
- Firm FEI
- 3002807099
- Firm city
- Perth
- Firm state
- Western Australia
- Firm country
- Australia
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28