Roche Molecular Systems, Inc.: cobas EGFR Mutation Test v2-Somatic gene mutation detection system Catalog Number :07248563190

Recall number
Z-0083-2022
Recalling firm
Roche Molecular Systems, Inc.
Product
cobas EGFR Mutation Test v2-Somatic gene mutation detection system Catalog Number :07248563190
Status
Open, Classified
Initiated
2021-09-03
Posted
not on file
Terminated
not on file
Reason
False Mutation Detected results for the exon 20 insertion (Ex20Ins) mutation when using the cobas EGFR Mutation Test v2, may lead to inappropriate treatment
Root cause
Under Investigation by firm
Action
Roche issued Urgent Medical Device Correction letter on 9/3/21 via UPS Ground. Letter states reason for recall, health risk and action to take: Follow the cobas DNA Sample Preparation Kit (catalog number 05985536190) IFU for sample input with regard to the number of FFPET sections and DNA concentration. Follow the cobas cfDNA Sample Preparation Kit (catalog number 07247737190) IFU for plasma samples. " If an Ex20Ins Mutation Detected result is generated with the cobas EGFR Mutation Test v2, you must confirm the result with another method (e.g., sequencing or other PCR-based tests). " Consider reviewing results generated since May 2021. Amivantamab was approved by the U.S. Food and Drug Administration (FDA) for NSCLC patients with EGFR Ex20Ins on May 21, 2021. Test result reports with Ex20Ins Mutation Detected may be considered for confirmatory testing using sequencing or other PCR-based tests, upon the discretion of a CAP/CLIA laboratory director. " Consult with your facility physician and/or pathologist to determine any clinical implications specific to your patients. " Complete all sections of the enclosed faxback form (TP-01381) and fax it to 1-855-208-4038 or email it to Roche8166@sedgwick.com. " File this UMDC for future reference. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-526-1247 if you have questions about the information contained in this UMDC
Quantity
507 US; 6,676 ex- US
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Canada, UK, Poland, Spain, Italy, Sweden, Greece, Denmark, Norway, Netherlands, France, Czech Republic, Belgium, Austria, Slovakia, Hungary, Switzerland, Lithuania, Slovenia. Serbia, North Macedonia, Myanmar, Lebanon, Russia, Croatia, Morocco, Kazakhstan, Dominican Republic, Israel, El Salvador, Vietnam, Malaysia, Taiwan, China, Japan, Hong Kong, Korea, Brazil, Philippines, Thailand, Chile, Colombia, Singapore, Argentina, Australia, India, Uruguay, Ecuador, Panama, Costa Rica, Pakistan, Peru, Indonesia, New Zealand.
Lot, serial or model codes
All lot numbers
510(k) numbers
not on file
PMA numbers
["P150047"]
Product code
OWD
Firm FEI
2243471
Firm city
S Branchburg
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register