Roche Molecular Systems, Inc.: cobas EGFR Mutation Test v2-Somatic gene mutation detection system Catalog Number :07248563190
- Recall number
- Z-0083-2022
- Recalling firm
- Roche Molecular Systems, Inc.
- Product
- cobas EGFR Mutation Test v2-Somatic gene mutation detection system Catalog Number :07248563190
- Status
- Open, Classified
- Initiated
- 2021-09-03
- Posted
- not on file
- Terminated
- not on file
- Reason
- False Mutation Detected results for the exon 20 insertion (Ex20Ins) mutation when using the cobas EGFR Mutation Test v2, may lead to inappropriate treatment
- Root cause
- Under Investigation by firm
- Action
- Roche issued Urgent Medical Device Correction letter on 9/3/21 via UPS Ground. Letter states reason for recall, health risk and action to take: Follow the cobas DNA Sample Preparation Kit (catalog number 05985536190) IFU for sample input with regard to the number of FFPET sections and DNA concentration. Follow the cobas cfDNA Sample Preparation Kit (catalog number 07247737190) IFU for plasma samples. " If an Ex20Ins Mutation Detected result is generated with the cobas EGFR Mutation Test v2, you must confirm the result with another method (e.g., sequencing or other PCR-based tests). " Consider reviewing results generated since May 2021. Amivantamab was approved by the U.S. Food and Drug Administration (FDA) for NSCLC patients with EGFR Ex20Ins on May 21, 2021. Test result reports with Ex20Ins Mutation Detected may be considered for confirmatory testing using sequencing or other PCR-based tests, upon the discretion of a CAP/CLIA laboratory director. " Consult with your facility physician and/or pathologist to determine any clinical implications specific to your patients. " Complete all sections of the enclosed faxback form (TP-01381) and fax it to 1-855-208-4038 or email it to Roche8166@sedgwick.com. " File this UMDC for future reference. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-526-1247 if you have questions about the information contained in this UMDC
- Quantity
- 507 US; 6,676 ex- US
- Distribution
- Worldwide distribution - US Nationwide distribution and the countries of Canada, UK, Poland, Spain, Italy, Sweden, Greece, Denmark, Norway, Netherlands, France, Czech Republic, Belgium, Austria, Slovakia, Hungary, Switzerland, Lithuania, Slovenia. Serbia, North Macedonia, Myanmar, Lebanon, Russia, Croatia, Morocco, Kazakhstan, Dominican Republic, Israel, El Salvador, Vietnam, Malaysia, Taiwan, China, Japan, Hong Kong, Korea, Brazil, Philippines, Thailand, Chile, Colombia, Singapore, Argentina, Australia, India, Uruguay, Ecuador, Panama, Costa Rica, Pakistan, Peru, Indonesia, New Zealand.
- Lot, serial or model codes
- All lot numbers
- 510(k) numbers
- not on file
- PMA numbers
- ["P150047"]
- Product code
- OWD
- Firm FEI
- 2243471
- Firm city
- S Branchburg
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28