Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Vascular, Inc.: Valiant Thoracic Stent Graft with Captivia Delivery System, Talent Thoracic Stent Graft with the Captivia Delivery System.

Recall number
Z-0083-2014
Recalling firm
Medtronic Vascular, Inc.
Product
Valiant Thoracic Stent Graft with Captivia Delivery System, Talent Thoracic Stent Graft with the Captivia Delivery System.
Status
Terminated
Initiated
2012-09-19
Posted
2013-10-25
Terminated
2013-10-25
Reason
Firm received reports of difficulty in flushing the graft cover of the delivery system during device preparation as required by IFU.
Root cause
Component design/selection
Action
A Medical Device Information letter was sent to all affected customers on September 19, 2012. Additionally, all Medtronic Field Representatives were notified of the issue via a Field Memo
Quantity
~1000
Distribution
International distribution:Austria, Belgium, Cyprus, Czech Republic, Denmark, Finland, Germany, Greece, Hungary, Iceland, Ireland, It lay, Latvia, Lithuania, Luxembourg, Netherlands, Norway, Poland, Portugal, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, United Kingdom.
Lot, serial or model codes
TBXXXXXXXXCP, TFXXXXXXXXCP, TWXXXXXXXXCP, TAXFXXXXXXXCP, TAXWXXXXXXXCP, VAMFXXXXXXXXTE, VAMCXXXXXXXXTE, VAMFXXXXXXXXTU, VAMCXXXXXXXXTU.
510(k) numbers
not on file
PMA numbers
["P100040/P070"]
Product code
MIH
Firm FEI
3005364322
Firm city
Santa Rosa
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register