Skip to the content

FDA Medical Device Recalls (openFDA)

Data Innovations, Inc.: Instrument Manager -Software Versions 8.05, 8.06, 8.07, 8.08, and 8.09. It is a stand-alone software package that provides connectivity between laboratory instruments and Laboratory Information Systems, aiding the flow of laboratory information, including test orders and results, specimen information, and patient information. Instrument Manager also provides data management functions that aid in the interpretation and management of laboratory results. The Instrument Manager software is put on a CD which is distributed in a box. User documentation consists of user manuals, PDF copies of which are on the CD Intended to be used by trained laboratory personnel as an aid in laboratory instrument interfacing and laboratory data management.

Recall number
Z-0083-2012
Recalling firm
Data Innovations, Inc.
Product
Instrument Manager -Software Versions 8.05, 8.06, 8.07, 8.08, and 8.09. It is a stand-alone software package that provides connectivity between laboratory instruments and Laboratory Information Systems, aiding the flow of laboratory information, including test orders and results, specimen information, and patient information. Instrument Manager also provides data management functions that aid in the interpretation and management of laboratory results. The Instrument Manager software is put on a CD which is distributed in a box. User documentation consists of user manuals, PDF copies of which are on the CD Intended to be used by trained laboratory personnel as an aid in laboratory instrument interfacing and laboratory data management.
Status
Terminated
Initiated
2009-05-11
Posted
2011-10-21
Terminated
2011-10-25
Reason
The "previous result" retrieved from the Specimen Management Database may not be the most recent "previous result".
Root cause
Software design
Action
Data Innovations sent an "IMPORTANT PRODUCT NOTICE" dated May 11, 2009 to all affected customers. The letter describes the product, problem, and actions to be taken by the customers. The letter offers temporary workarounds for customers to use until corrected versions of the software is complete. Versions of the corrections will be provided on a staggered schedule. Customers will be notified by e-mail. Questions regarding this notice are directed to Support at 802-658-1955.
Quantity
188
Distribution
USA (nationwide) and the country of Canada.
Lot, serial or model codes
S/W Versions 8.05, 8.06, 8.07, 8.08 and 8.09.
510(k) numbers
["BK080052"]
PMA numbers
not on file
Product code
JQP
Firm FEI
3001451449
Firm city
South Burlington
Firm state
VT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register