Invivo Corporation: Xper Flex Cardio PhysioMonitoring system The Xper Flex Cardio physiomonitoring system is used to facilitate invasive investigation of heart and vascular disease when non-invasive indicators warrant such. Its users, responsible to interpret the data made available, will be professional health care providers, such as physicians, registered nurses, cardiovascular technicians, radiology technicians, and other clinical health care providers. The Xper Flex Cardio physiomonitoring system may be used to display and analyze surface ECG (electrocardiogram), respiration, invasive pressure, Sp02 (Pulse Oximetry), End Tidal C02 (carbon dioxide) and non-invasive pressure waveforms; surface body temperature and thermal cardiac output curves. The system is capable of processing/analyzing information such as multi-channel ECG signals and displaying a graphical ST segment map. The Xper Flex Cardio physiomonitoring system makes measurements that assist physicians and other clinicians to evaluate a patient's overall cardiopulmonary performance, conductive system and general vascular status.

Recall number
Z-0082-2018
Recalling firm
Invivo Corporation
Product
Xper Flex Cardio PhysioMonitoring system The Xper Flex Cardio physiomonitoring system is used to facilitate invasive investigation of heart and vascular disease when non-invasive indicators warrant such. Its users, responsible to interpret the data made available, will be professional health care providers, such as physicians, registered nurses, cardiovascular technicians, radiology technicians, and other clinical health care providers. The Xper Flex Cardio physiomonitoring system may be used to display and analyze surface ECG (electrocardiogram), respiration, invasive pressure, Sp02 (Pulse Oximetry), End Tidal C02 (carbon dioxide) and non-invasive pressure waveforms; surface body temperature and thermal cardiac output curves. The system is capable of processing/analyzing information such as multi-channel ECG signals and displaying a graphical ST segment map. The Xper Flex Cardio physiomonitoring system makes measurements that assist physicians and other clinicians to evaluate a patient's overall cardiopulmonary performance, conductive system and general vascular status.
Status
Terminated
Initiated
2017-08-11
Posted
not on file
Terminated
2024-09-20
Reason
Intermittent communication between the host system and the Flex Cardio
Root cause
Software Design Change
Action
On August 2017 the firm sent letters to their consignees indicating the following: This document contains important information for the continued safe and proper use of your equipment Please review the following information with all members of your staff who need to be aware of the contents of this communication. It is important to understand the implications of this communication. Please retain a copy with the equipment Instruction for Use. ------------------------------------------- A problem has been detected in the Philips Xper Flex Cardio Patient Monitoring System (Flex Cardio) that, if it were to recur, could pose a risk for patients or users. This Field Safety Notice is intended to inform you about: " what the problem is and under what circumstances it can occur " the actions that should be taken by the customer / user in order to prevent risks for patients or users " the actions planned by Philips to correct the problem. Because of intermittent communication between the host system and the Flex Cardio, it is possible for any of the following conditions to occur: " Inability of the Flex Cardio to connect to the host system. " Boom Monitor (display) may not display all active waveform and/or vital sign data. " Delayed audible and visual alarms at the Flex Cardio and Boom Monitor respectively, due to a mismatch between the alarm limits on the host system and the Flex Cardio/Boom Monitor. " After patient admission and case initiation at the host system, the case may not be fully transmitted to the Flex Cardio. In this case, the Boom Monitor in the exam/procedure room will display vital sign data, but the Flex Cardio will not provide audible alarms. " After patient discharge and case termination at the host system, case termination may not be transmitted to the Flex Cardio. In this case, the Boom Monitor in the exam/procedure room will continue to display vital sign data, and the Flex Cardio will continue to provide audible alarms. Our re
Quantity
529
Distribution
Worldwide Distribution - US including AK, AL, AZ, CA, CO, DE, FL, GA, IA, ID, IL, KY, LA, MA, MI, MO, NC, NE, NM, NY, OH, OR, OK, PA, SD, TN, TX, VA, WA, WV and Internationally to Germany, Hungary, India, Netherlands, Russia, Saudi Arabia, South Africa and Tunisia.
Lot, serial or model codes
Model 453564621791 and 453564621801 - ALL Flex Cardio Devices, Revision C
510(k) numbers
["K101571"]
PMA numbers
not on file
Product code
MWI
Firm FEI
1051786
Firm city
Orlando
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register