FDA Medical Device Recalls (openFDA)
Maquet Medical Systems USA: MAQUET Getinge Group CARDIOHELP Base Unit REF 70104.7999 REF 70104.8012 Manufactured for: MAQUET Cardiopulmonary AG Made in Germany. Cardiopulmonary support system.
- Recall number
- Z-0082-2014
- Recalling firm
- Maquet Medical Systems USA
- Product
- MAQUET Getinge Group CARDIOHELP Base Unit REF 70104.7999 REF 70104.8012 Manufactured for: MAQUET Cardiopulmonary AG Made in Germany. Cardiopulmonary support system.
- Status
- Terminated
- Initiated
- 2013-09-12
- Posted
- 2013-10-25
- Terminated
- 2017-06-07
- Reason
- It has come to the attention of MAQUET that in certain rare instances, the CARDIOHELP-i device may display an error message with audible alarm, indicating "battery defective." The alarm will occur-upon start up or when the main power supply is unplugged from the main power source. The human machine interface touch screen of the CARDIOHELP-i device may gradually become black due to the human machine interface (HMI) back light shutting down.
- Root cause
- Software Manufacturing/Software Deployment
- Action
- MAQUET issued their customers "Urgent-Medical Device Field Correction" notices/Customer Fax Response forms dated September 16, 2013 via FedEx on September 16, 2013. The notification informs the customer of the issue with the product; how to identified affected product; and the action needed to be taken by the customer. The field correction includes service on the affected product. Authorized MAQUET Service Technicians will service the affected product to correct the issue. Customers were requested to complete and fax back (or email a scanned copy) of the enclosed response form. Customers could contact their local field representative or the MAQUET Customer Service line at 1.888.627.8383 (press option 2 followed by option 2) Monday through Friday, 8am to 6pm EST. For questions regarding this recall call 973-709-7000.
- Quantity
- 63
- Distribution
- Nationwide Distribution including AZ, FL, CO, TX, OH, VA, MN, WI, SC, IL, ND, UT, CA, OR, NY, IA, CT, SD, MI, HI, AL, MD, KS, MO
- Lot, serial or model codes
- Part No. 70104.8012 Serial Nos. 90410440; 90410452; 90410455; 90410456; 90410457; 90410459; 90410469; 90410474; 90410480; 904104781; 90410482; 90410484; 90410498; 90410499; 90410500; 90410501; 90410502; 90410503; 90410521; 90410522; 90410524; 90410526; 90410527; 90410531; 90410533; 90410534; 90410535; 90410550; 90410551; 90410552; 90410568; 90410571; 90410572; 90410573; 90410575; 90410576; 90410577; 90410580; 90410581; 90410582; 90410584; 90410585; 90410592; 90410596; 90410599; 90410600; 90410605; 90410606; 90410607; 90410608; 90410609; 90410610; 90410611; 90410612; 90410626; 90410638; 90410640; 90410642; 90410644; 90410646; 90410647; 90410650; 90410651; 90410652; 90410653; 90410656; 90410659; 90410660; 90410661; 90410662; 90410663; 90410664; 90410665; 90410666; 90410667; 90410668; 90410670; 90410700; 90410708; 90410846; 90410750; 90410786
- 510(k) numbers
- ["K102726"]
- PMA numbers
- not on file
- Product code
- DTQ
- Firm FEI
- 3008355164
- Firm city
- Wayne
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28