Biomet, Inc.: Biomet Regenerex Primary Taper Cap Item Number 141269

Recall number
Z-0081-2021
Recalling firm
Biomet, Inc.
Product
Biomet Regenerex Primary Taper Cap Item Number 141269
Status
Terminated
Initiated
2020-09-04
Posted
not on file
Terminated
2021-08-31
Reason
Product not properly being aligned with the adequate gamma sterilization group and result in the product not being properly sterilized
Root cause
Process change control
Action
Zimmer Biomet notified on 9/2/20 Distributors with direct sales and/or fielded inventory via email and traceable letter sent via FedEx. Hospitals with direct sales or current inventory, as well as hospitals that have consumed products in the past 12 months will be notified via FedEx. Surgeons that have consumed products with in the past 12 months notified via FedEx. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form will be returned even if you do not have affected products at your facility. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. 3. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5. If you ha
Quantity
US: 62 units; OUS: 176 units
Distribution
Nationwide Foreign: Canada, FINLAND, NETHERLANDS, UNITED KINGDOM
Lot, serial or model codes
Lot Numbers: 015800 324440 685300 023900 324440R 696280 058930 389060 821980 200720 418430 890320 313260 646040 997610
510(k) numbers
["K080361"]
PMA numbers
not on file
Product code
MBH
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register