Ortho-Clinical Diagnostics: VITROS 5600 Integrated System-For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, Catalog Number: 6802413

Recall number
Z-0079-2019
Recalling firm
Ortho-Clinical Diagnostics
Product
VITROS 5600 Integrated System-For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, Catalog Number: 6802413
Status
Terminated
Initiated
2018-08-15
Posted
not on file
Terminated
2022-01-10
Reason
Potential for the VITROS 5600 Integrated System to dispense an insufficient sample volume onto a MicroSlide without suppressing the potentially incorrect result.
Root cause
Software design
Action
ORTHO PLUSSM is e-Communications, a service that provides subscribing customers with Ortho Clinical Diagnostics (Ortho) product notifications electronically on 15 August 2018, a customer letter (Ref. CL2018-174) was sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) to all customers who received VITROS 5600 Integrated Systems to inform them of the potential for the VITROS 5600 Integrated System to dispense an insufficient sample volume onto a MicroSlide without suppressing the potentially incorrect result. Ortho also provided specific actions for customers to take when they observe unexpected MicroSlide assay results for a sample associated with TE1-504 and/or TE1-594 condition codes. Customers were also instructed to discuss any concerns related to this issue and previously reported results with the Laboratory Medical Director.
Quantity
U.S.= 1255 ex-U.S.= 1504
Distribution
Nationwide Foreign: France, Brazil, Chile, Great Britain, Phillippines, Spain
Lot, serial or model codes
Serial Numbers : J56000024 through J56003300
510(k) numbers
["K081543"]
PMA numbers
not on file
Product code
JLW
Firm FEI
1000136573
Firm city
Rochester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register