FDA Medical Device Recalls (openFDA)
Medtronic Perfusion Systems: Medtronic Intersept Custom Tubing Pack, which are designed as specified by the user. The following models contain a Pressure Display Set. Sterilized by Ethylene Oxide. Non-pyrogenic. Manufacturer: Medtronic, Inc. 710 Medtronic Parkway, Minneapolis, MN 55432-5604. Product Usage: This product is intended for use in monitoring catheter, cannula, or line pressures associated with cardiopulmonary bypass equipment and/or related products.
- Recall number
- Z-0079-2012
- Recalling firm
- Medtronic Perfusion Systems
- Product
- Medtronic Intersept Custom Tubing Pack, which are designed as specified by the user. The following models contain a Pressure Display Set. Sterilized by Ethylene Oxide. Non-pyrogenic. Manufacturer: Medtronic, Inc. 710 Medtronic Parkway, Minneapolis, MN 55432-5604. Product Usage: This product is intended for use in monitoring catheter, cannula, or line pressures associated with cardiopulmonary bypass equipment and/or related products.
- Status
- Terminated
- Initiated
- 2011-09-09
- Posted
- 2011-10-20
- Terminated
- 2012-10-11
- Reason
- Pressure Display Set (models 61000, 61006, 62000, 62006) DFU will be updated to clarify that the intended use of the disposables is only for cardiopulmonary bypass surgery (procedures up to 6 hours). Use beyond 6 hours may result in air leakage in the system resulting in a pressure reading that is not representative of the actual line pressure.
- Root cause
- Device Design
- Action
- Medtronic Perfusion Systems, sent an "Urgent Medical Device Notification letter dated September 9, 2011, via UPS 2-day delivery to all affected customers. The letter identified the affected devices, described the issues, and the actions needed to be taken. The letter indicates when approved, the updated Directions for Use for the Pressure Display Box and Pressure Display Sets will be sent to customers. Customers were asked to complete the Customer Confirmation Certificate and return to Medtronic per the instructions on the form. The letter states that if the recalled products have been further distributed, forward this letter immediately to all customers who received the affected products, and gather proof of notification. For further questions concerning this notification, contact Medtronic Cardiovascular Technical Support at 877-526-7890 or Medtronic Sales Representative.
- Quantity
- 74,804
- Distribution
- Worldwide Distribution - USA (nationwide) including Puerto Rico, and Virgin Islands (U.S.) and the countries of: Argentina, Australia, Austria, Belarus, Belgium, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, El Salvador, France, Georgia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Israel, Italy, Jamaica, Japan, Korea, Kuwait, Latvia, Lebanon, Libya, Malaysia, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Poland, Portugal, Romania, Saudi Arabia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, United Arab Emirates, United Kingdom, Venezuela, and Yemen.
- Lot, serial or model codes
- Models 0K57R2, 0K82R15, 0M63R7, 0M68R6, 0N41R14, 0T50R5, 0V67R27, 0X52R3, 0Y35R, 0Y95R30, 0Z30R11, 1D41R16, 1F66R12, 1G59R1, 1J74R4, 1M90R8, 1P09R2, 2031R7, 2C88R8, 2C89R7, 2C90R7, 2C91R5, 2D61R2, 2D64R16, 2H08R3, 2K85R1, 2K87R1, 2K88R1, 2N90R9, 2T28R1, 3B74R2, 3T73R1, 3W29R, 4B42R5, 4C03R, 4D82R3, 4M55R, 4N04R, 4R43R, 5788R4, 5A45R3, 5B51R9, 5B94R, 5C33R2, 5C45R3, 5C92R9, 5D83R2, 5F68R1, 5G28R4, 5H65R, 5J00R2, 5P62R, 5W13R6, 5Z18R, 6G05R, 6J55R1, 6J88R4, 6J95R8, 6N12R, 6Q49R1, 6Q49R1, 6R08R, 6R09R, 6R44R2, 6S66R, 6S75R1, 6V25R, 6Y04R5, 6Y07R6, 6Y59R3, 7A69R4, 7A70R1, 7B56R1, CB0L44R16, CB0M02R5, CB0M79R8, CB1Q18R11, CB1S33R9, CB2D32R5, CB2W45R7, CB2W86R4, CB3K45R9, CB3K91R1, CB3M40R17, CB3Y46R9, CB3Y47R10, CB4800R18, CB4953R16, CB4958R16, CB4B68R8, CB4E78R7, CB4F45R15, CB4K64R6, CB4L83R4, CB4M60R2, CB4S43R9, CB4W10R10, CB4X84R, CB4Y71R14, CB4Z87R6, CB5A30R10, CB5F84R12, CB5J33R6, CB5L15R13, CB5L79R, CB5L90R16, CB5Q71R2, CB5Q74R1, CB5R04R1, CB5R46R5, CB5U81R, CB6A80R6, CB6G02R4, CB6H12R, CB6H45R4, CB6H76R, CB6H77R, CB6H78R, CB6J66R2, CB6J68R3, CB6J75R3, CB6L12R3, CB6L13R, CB6L14R, CB6L60R2, CB6M11R1, CB6Q15R, CB6Q16R1, CB6Q62R, CB6Q78R2, CB6S17R, CB6U12R, CB6U76R1, CB6X94R2, CB6X98R3, CB6Y02R6, CB6Y34R2, CB7C31R, DLP2X17R, HY0F32R17, HY2280R16, HY2U14R2, HY3U59R12, HY4A96R9, HY4V52R4, HY4V82R3, HY5D07R2, HY6B26R2, HY6G55R6, HY6G56R6, HY6H03R5, HY6H03R5, HY6J33R3, HY6J99R, HY6R36R1, HY6R36R1, HY6U70R2, HY6V34R2, HY6Z04R1, M331105H, M331105K, M336902L, M392810D, M392810E, M392811C, M392811D, M392811E, M440115G, M440115H, M440115J, M440115K, M440115L, M440129A, M440130A, M441002E, M443824F, M443825E, M492609A, M499301B, M530710A, M530710C, M530710D, M960208A, M960454B, P390806A, S330115A, SS6X67R1, SS6Z09R1, SSHY6E81R2, TL2H89R7, TL3Y45R1, TL3Z67R4, TL4E64R, TL4E66R10, TL4E66R10, TL5398R21, TL5C09R4, TL5M01R7, TL5R15R4, TL6A91R2, TL6C91R2, TL6D87R2, TL6H39R5, TL6W08R3, TL6X29R4, TL6X36R2, TL6X62R5, TL6Y92R, TL7A15R2.
- 510(k) numbers
- ["K852232"]
- PMA numbers
- not on file
- Product code
- DXS
- Firm FEI
- 1000116158
- Firm city
- Brooklyn Park
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28