Intuitive Surgical, Inc.: Intuitive Surgical, daVinci S Surgical System, Model Number: IS2000, A5.1 P1 and A5.0 P5 software level, Surgical Endoscopic Instrument Control System, Intuitive Surgical, Sunnyvale, CA 94086.

Recall number
Z-0079-2008
Recalling firm
Intuitive Surgical, Inc.
Product
Intuitive Surgical, daVinci S Surgical System, Model Number: IS2000, A5.1 P1 and A5.0 P5 software level, Surgical Endoscopic Instrument Control System, Intuitive Surgical, Sunnyvale, CA 94086.
Status
Terminated
Initiated
2007-03-30
Posted
2008-02-21
Terminated
2008-09-05
Reason
System Lock-Up: Software anomalies could cause product failure during use, or on start-up. System transitions to a safe "soft-lock" state.
Root cause
Software design
Action
Intuitive Field Service Engineers installed a software upgrade beginning on 03/30/2007, and left a notification letter that the upgrade had been performed at each site. New software level is A5.1 P2.
Quantity
159
Distribution
Worldwide Distribution - USA, Belgium, China, France, India, Italy, Japan, Korea, Spain, Sweden, Switzerland, UK and Mexico.
Lot, serial or model codes
Model Number: IS 2000, Software Revision Numbers: A5.0 P5 and A5.1 P1. System Numbers: all starting with "SG"-- 043, 156, 088, 128, 065, 104, 106, 132, 019, 051, 058, 083, 092, 096, 165, 170, 098, 105, 157, 050, 112, 125, 138, 149, 034, 061, 121, 030, 044, 052, 054, 068, 069, 123, 147, 151, 013, 113, 152, 014, 049, 077, 134, 136, 066, 085, 091, 129, 087, 139, 140, 042, 047, 031, 089, 017, 021, 022, 115, 116, 137, 155, 168, 169, 041, 095, 119, 163, 164, 114, 057, 060, 101, 117, 122, 029, 020, 028, 056, 093, 094, 141, 103, 072, 084, 011, 024, 036, 053, 059, 144, 145, 146, 026, 074, 154, 018, 027, 055, 070, 109, 110, 111, 118, 120, 135, 099, 232, 033, 082, 148, 012, 015, 039, 067, 102, 124, 161, 166, 037, 097, 040, 064, 090, 127, 133, 025, 062, 143, 160, 038, 153, 035, 071, 073, 107, 158, 167, 086, 131, 063, 080, 078, 159, 076, 045, 081, 108, 150, 046, 075, 130, 004, 005, 006, 007, 010, 048, and 126
510(k) numbers
["K050369","K060322"]
PMA numbers
not on file
Product code
NAY
Firm FEI
3001675293
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register