Intuitive Surgical, Inc.: Intuitive Surgical, daVinci S Surgical System, Model Number: IS2000, A5.1 P1 and A5.0 P5 software level, Surgical Endoscopic Instrument Control System, Intuitive Surgical, Sunnyvale, CA 94086.
- Recall number
- Z-0079-2008
- Recalling firm
- Intuitive Surgical, Inc.
- Product
- Intuitive Surgical, daVinci S Surgical System, Model Number: IS2000, A5.1 P1 and A5.0 P5 software level, Surgical Endoscopic Instrument Control System, Intuitive Surgical, Sunnyvale, CA 94086.
- Status
- Terminated
- Initiated
- 2007-03-30
- Posted
- 2008-02-21
- Terminated
- 2008-09-05
- Reason
- System Lock-Up: Software anomalies could cause product failure during use, or on start-up. System transitions to a safe "soft-lock" state.
- Root cause
- Software design
- Action
- Intuitive Field Service Engineers installed a software upgrade beginning on 03/30/2007, and left a notification letter that the upgrade had been performed at each site. New software level is A5.1 P2.
- Quantity
- 159
- Distribution
- Worldwide Distribution - USA, Belgium, China, France, India, Italy, Japan, Korea, Spain, Sweden, Switzerland, UK and Mexico.
- Lot, serial or model codes
- Model Number: IS 2000, Software Revision Numbers: A5.0 P5 and A5.1 P1. System Numbers: all starting with "SG"-- 043, 156, 088, 128, 065, 104, 106, 132, 019, 051, 058, 083, 092, 096, 165, 170, 098, 105, 157, 050, 112, 125, 138, 149, 034, 061, 121, 030, 044, 052, 054, 068, 069, 123, 147, 151, 013, 113, 152, 014, 049, 077, 134, 136, 066, 085, 091, 129, 087, 139, 140, 042, 047, 031, 089, 017, 021, 022, 115, 116, 137, 155, 168, 169, 041, 095, 119, 163, 164, 114, 057, 060, 101, 117, 122, 029, 020, 028, 056, 093, 094, 141, 103, 072, 084, 011, 024, 036, 053, 059, 144, 145, 146, 026, 074, 154, 018, 027, 055, 070, 109, 110, 111, 118, 120, 135, 099, 232, 033, 082, 148, 012, 015, 039, 067, 102, 124, 161, 166, 037, 097, 040, 064, 090, 127, 133, 025, 062, 143, 160, 038, 153, 035, 071, 073, 107, 158, 167, 086, 131, 063, 080, 078, 159, 076, 045, 081, 108, 150, 046, 075, 130, 004, 005, 006, 007, 010, 048, and 126
- 510(k) numbers
- ["K050369","K060322"]
- PMA numbers
- not on file
- Product code
- NAY
- Firm FEI
- 3001675293
- Firm city
- Sunnyvale
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28