Skip to the content

FDA Medical Device Recalls (openFDA)

Smiths Medical ASD Inc.: Medfusion Syringe Pump, Model 3500-0600-XX: a) 0600-00; b) 0600-01; c) 0600-50; d) 0600-51; e) 0600-82; f) 0600-249; software versions: v3, v4, v5, and v6

Recall number
Z-0078-2024
Recalling firm
Smiths Medical ASD Inc.
Product
Medfusion Syringe Pump, Model 3500-0600-XX: a) 0600-00; b) 0600-01; c) 0600-50; d) 0600-51; e) 0600-82; f) 0600-249; software versions: v3, v4, v5, and v6
Status
Open, Classified
Initiated
2023-08-16
Posted
2023-10-06
Terminated
not on file
Reason
A force sensor in the occlusion detector may drift out of calibration leading to increased occlusion detection times, false occlusion alarms, or a System Failure Alarm. If the force sensor calibration shift is large enough, the pump will display a System Failure Alarm (including Force Sensor BGND Test, Force Sensor Bridge Test, or Force Sensor Test). However, if the calibration shift is not large enough to trigger a System Failure Alarm, the threshold to detect an occlusion may increase, increasing the time to occlusion detection, or the threshold may decrease, leading to false occlusion alarms. Although shifts in the force sensor calibration may occur over time with any device, an increased potential for such shifts has been reported in devices produced before April 2022 due to mechanical interference between parts of the plunger head assembly. Out of an abundance of caution, we are notifying all customers of this potential issue.
Root cause
Under Investigation by firm
Action
Smiths Medical sent an URGENT MEDICAL DEVICE CORRECTION notice to its direct consignees on 08/16/2023 by traceable mailing. The notice explained the issue, potential risk to the patient, and requested the following actions be taken: Customer Required Actions 1. Locate all Medfusion Model 3500 and Medfusion Model 4000 syringe pumps in your possession and ensure all users or potential users of these devices are immediately made aware of this notification and proposed mitigations. 2. Ensure all Medfusion Model 3500 and Medfusion Model 4000 syringe pumps in your possession have undergone all tests in the Annual Maintenance List including the new Force Sensor Check within the past 12 months, including new pumps that have not yet undergone annual maintenance. 3. Complete and return the attached Response Form to smithsmedical6995@sedgwick.com within ten days of receipt to acknowledge your understanding of this notification. 4. DISTRIBUTORS: If you have distributed potentially affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to smithsmedical6995@sedgwick.com
Quantity
20383 pumps
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Lot, serial or model codes
a) 0600-00, UDI/DI 10610586041259; b) 0600-01, UDI/DI 10610586041242; c) 0600-50, UDI/DI 10610586043284; d) 0600-51, UDI/DI 10610586043291; e) 0600-82, UDI/DI 10610586043017; f) 0600-249, UDI/DI 10610586041297; All serial numbers
510(k) numbers
["K040899"]
PMA numbers
not on file
Product code
FRN
Firm FEI
3012307300
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register