Skip to the content

FDA Medical Device Recalls (openFDA)

Teleflex Medical: Hem-o-lok Large Polymer Ligating Clips, REF #544240, Rx only, Sterile, Teleflex Medical, Research Triangle Park, NC 27709. Teleflex Medicals Hem-o-lok polymer ligating clips are single use, sterile, non-absorbable medial devices that have been designed to ligate vessels and tissue structures during various types of surgical procedures. Vessels and tissue structures in the range of 0.5 to 16 mm can be effectively ligated with Hem-o-lok clips.

Recall number
Z-0078-2013
Recalling firm
Teleflex Medical
Product
Hem-o-lok Large Polymer Ligating Clips, REF #544240, Rx only, Sterile, Teleflex Medical, Research Triangle Park, NC 27709. Teleflex Medicals Hem-o-lok polymer ligating clips are single use, sterile, non-absorbable medial devices that have been designed to ligate vessels and tissue structures during various types of surgical procedures. Vessels and tissue structures in the range of 0.5 to 16 mm can be effectively ligated with Hem-o-lok clips.
Status
Terminated
Initiated
2012-09-25
Posted
2012-10-18
Terminated
2012-11-08
Reason
Product is being recalled due to the possibility that the tray may contain pinholes, compromising product sterility.
Root cause
Packaging process control
Action
Teleflex sent a Urgent Medical Device Recall Notification dated September 25, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to check their inventory and quarantine any affected inventory found. They were to complete the enclosed Field Corrective Action Acknowledgement Form and return it to the Teleflex Medical sales representative who will then remove the affected product from the facility and ensure its return to Teleflex Medical. We sincerely apologize for any inconvenience this action may cause your operations. If you have any questions please contact your local sales representative or Customer Service at 1-866-246-6990.
Quantity
126 units
Distribution
Nationwide Distribution including the state of Nevada.
Lot, serial or model codes
Catalog No. 544240, Lot #01E1200585
510(k) numbers
["K993157"]
PMA numbers
not on file
Product code
FZP
Firm FEI
3005747797
Firm city
Research Triangle Park
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register