Skip to the content

FDA Medical Device Recalls (openFDA)

Medica Corporation: EasyRA Cuvette Segment 1, P/N 007000-001 (In Vitro Diagnostic ) EasyRA Cuvette Segments included in the following Kits: EasyRA Cuvettes, REF 10700-4 Startup Kit - Chemistry, REF 10990 Startup Kit - Chemistry & ISE, REF 10991

Recall number
Z-0078-2011
Recalling firm
Medica Corporation
Product
EasyRA Cuvette Segment 1, P/N 007000-001 (In Vitro Diagnostic ) EasyRA Cuvette Segments included in the following Kits: EasyRA Cuvettes, REF 10700-4 Startup Kit - Chemistry, REF 10990 Startup Kit - Chemistry & ISE, REF 10991
Status
Terminated
Initiated
2010-03-29
Posted
2010-10-18
Terminated
2011-06-07
Reason
Chemistry analyzers may report Calcium results below the sample's actual concentration. Reported results may be ten to fifteen percent lower than expected sample results. The defect is a shortened path length at specific cuvette positions of those segments manufactured using molding cavity 1. The shortened path length was determined by measuring the cuvette dimensions with a Co-ordinate Measuri
Root cause
Other
Action
Medica issued an Urgent Product Recall letter dated March 29, 2010 to customers. Users were requested to cease using and discard the affected product, and complete and return a Response Form in order to obtain replacement product. Customers can contact Medica Technical Support at 877-777-5895.
Quantity
14,926
Distribution
Worldwide Distribution: USA, and the countries of Bulgaria, Canada, India, Mexico, Panama, Phillipines, Russia, Turkey, New Zealand, South Africa, United Arab Emirates, and Venezuela.
Lot, serial or model codes
Lot Numbers:  2009/A2810, 2009/A2910, 2009/A3010, 2010/A2001, 2010/A21 01, 2010/A1901, 2009/A0211, 2009/A0311
510(k) numbers
["K070057","K072249"]
PMA numbers
not on file
Product code
JJE
Firm FEI
1220972
Firm city
Bedford
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register