FDA Medical Device Recalls (openFDA)
Medtronic Perfusion Systems: Medtronic Pressure Display Box, Models 66000, 64000, 63000, 60000. It is capable of measuring pressure between -100 mmHg and +500 mmHg with an accuracy of +/-5 mmHg from -100 mmHg to +300 mmHg and +/-10 mmHg from +301 to +500 mmHg. The Pressure Display Box has an upper and lower alarm setting. The alarm will be activated when the pressure is equal to or beyond the set value. Manufacturer: Medtronic, Inc. 710 Medtronic Parkway, Minneapolis, MN 55431-5604. Product Usage: This product is intended for use in displaying line pressures during cardiopulmonary bypass surgery.
- Recall number
- Z-0077-2012
- Recalling firm
- Medtronic Perfusion Systems
- Product
- Medtronic Pressure Display Box, Models 66000, 64000, 63000, 60000. It is capable of measuring pressure between -100 mmHg and +500 mmHg with an accuracy of +/-5 mmHg from -100 mmHg to +300 mmHg and +/-10 mmHg from +301 to +500 mmHg. The Pressure Display Box has an upper and lower alarm setting. The alarm will be activated when the pressure is equal to or beyond the set value. Manufacturer: Medtronic, Inc. 710 Medtronic Parkway, Minneapolis, MN 55431-5604. Product Usage: This product is intended for use in displaying line pressures during cardiopulmonary bypass surgery.
- Status
- Terminated
- Initiated
- 2011-09-09
- Posted
- 2011-10-20
- Terminated
- 2012-10-11
- Reason
- The DFU for the Medtronic Pressure Display Box, model 66000, states that, for display readings of -100 to +300 mmHg there will be an accuracy of +/-5 mmHg, and for display readings of +301 to +500 mmHg there will be an accuracy of +/-10 mmHg. Through reports from customers, Medtronic has become aware that the Pressure Display Box may not always accurately display readings. During use in the -40
- Root cause
- Device Design
- Action
- Medtronic Perfusion Systems, sent an "Urgent Medical Device Notification letter dated September 9, 2011, via UPS 2-day delivery to all affected customers. The letter identified the affected devices, described the issues, and the actions needed to be taken. The letter indicates when approved, the updated Directions for Use for the Pressure Display Box and Pressure Display Sets will be sent to customers. Customers were asked to complete the Customer Confirmation Certificate and return to Medtronic per the instructions on the form. The letter states that if the recalled products have been further distributed, forward this letter immediately to all customers who received the affected products, and gather proof of notification. For further questions concerning this notification, contact Medtronic Cardiovascular Technical Support at 877-526-7890 or Medtronic Sales Representative.
- Quantity
- 1694 units
- Distribution
- Worldwide Distribution - USA (nationwide) including Puerto Rico, and Virgin Islands (U.S.) and the countries of: Argentina, Australia, Austria, Belarus, Belgium, Bulgaria, Canada, Chile, China, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, El Salvador, France, Georgia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Israel, Italy, Jamaica, Japan, Korea, Kuwait, Latvia, Lebanon, Libya, Malaysia, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Poland, Portugal, Romania, Saudi Arabia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, United Arab Emirates, United Kingdom, Venezuela, and Yemen.
- Lot, serial or model codes
- serial numbers: 301010, 301012, 301013, 301015, 301016, 301017, 301018, 301019, 301020, 301023, 301024, 301025, 301026, 301027, 301029 thru 301041, 301043 thru 301050, 301052, 301053, 301054, 301055, 301056, 301057, 301058, 301060, 301061, 301062, 301063, 301068 thru 301085, 301087, 301088, 301090, 301091, 301092, 301093, 301094, 301300, 301301, 301302, 301303, 301371 thru 301383, 301385, 301387, 301388, 301389, 301390, 301394, 301396, 301397, 301399 thru 301407, 301409, 301410, 301431 thru 301450, 301460, 301461, 301462, 301464, 301467 thru 301485, 301487 thru 301507, 301509 thru 301526, 301560 thru 301663, 301665 thru 301704, 301707 thru 301751, 301753, 301754, 301755, 301756, 301758 thru 301817, 301819 thru 301842, 301844 thru 302148, 302150 thru 302161, PDBR400000 thru PDBR400127, PDBR400129 thru PDBR400835, PDBR400847 thru PDBR401104, PDBR401107 thru PDBR401178.
- 510(k) numbers
- ["K852232"]
- PMA numbers
- not on file
- Product code
- DXS
- Firm FEI
- 1000116158
- Firm city
- Brooklyn Park
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28