Skip to the content

FDA Medical Device Recalls (openFDA)

Angiodynamics Inc. (Navilyst Medical Inc.): BioFlo Port with ENDEXO and PASV Technology-AngioDynamics BIOFLO 8F SINGLE PLASTIC NON-FILLED; VALVED PG UPN: H965440210 Catalog Number: 44-021 Box of 1 Port - Product Usage - Long-term access to the central venous system for administration of fluids including but not limited to hydration fluids, chemotherapy, analgesics, nutritional therapy and blood products. The device is also indicated for blood specimen withdrawal.

Recall number
Z-0076-2019
Recalling firm
Angiodynamics Inc. (Navilyst Medical Inc.)
Product
BioFlo Port with ENDEXO and PASV Technology-AngioDynamics BIOFLO 8F SINGLE PLASTIC NON-FILLED; VALVED PG UPN: H965440210 Catalog Number: 44-021 Box of 1 Port - Product Usage - Long-term access to the central venous system for administration of fluids including but not limited to hydration fluids, chemotherapy, analgesics, nutritional therapy and blood products. The device is also indicated for blood specimen withdrawal.
Status
Terminated
Initiated
2018-07-24
Posted
not on file
Terminated
2019-05-17
Reason
BioFlo Ports was partially manufactured with the incorrect sterile port assembly. The actual sterile port assembly used was, 8F SL Xcela Plus Titanium Valved Non-Filled, when it should have been, 8F SL BioFlo Plastic Valved Non-Filled
Root cause
Unknown/Undetermined by firm
Action
On July 24, 2018 AngioDynamics initiated this recall action by notifying the affected consignee via telephone call.Per the telephone conversation, the consignee was instructed to segregate the affected device and quarantine the product until further direction from AngioDynamics. A follow-up letter issued dated 7/24/18 email response form to be completed and returned.
Quantity
1 box
Distribution
US Nationwide in the state of IN.
Lot, serial or model codes
Lot Number: 5363172
510(k) numbers
["K131694"]
PMA numbers
not on file
Product code
LJT
Firm FEI
3017892510
Firm city
Glens Falls
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register