Skip to the content

FDA Medical Device Recalls (openFDA)

Sorin Group USA, Inc.: FlexFlow Venous Cannula Model/Catalog Number 200-200 The FlexFlow Venous Cannula is a single use sterile device. It is a size 23 French with an outside diameter of 0.30 inch (7.7 mm) and an effective working length of 11.2 inch (28.5 cm). The FlexFlow Venous Cannula is an open lumen polymer tube incorporating wire reinforcement in distal sections. The distal sections of the cannula are perforated with multiple openings to allow increased fluid flow. The clear proximal section is not reinforced to allow clamping the proximal end that terminates in a 3/8 inch (9.5 mm) barbed connector for standard cardiopulmonary bypass tubing. Each cannula is furnished with a mated obturator. Each obturator has a malleable wire. Each component is packaged inside a protective sheath in a single sterile, sealed pouch

Recall number
Z-0076-2018
Recalling firm
Sorin Group USA, Inc.
Product
FlexFlow Venous Cannula Model/Catalog Number 200-200 The FlexFlow Venous Cannula is a single use sterile device. It is a size 23 French with an outside diameter of 0.30 inch (7.7 mm) and an effective working length of 11.2 inch (28.5 cm). The FlexFlow Venous Cannula is an open lumen polymer tube incorporating wire reinforcement in distal sections. The distal sections of the cannula are perforated with multiple openings to allow increased fluid flow. The clear proximal section is not reinforced to allow clamping the proximal end that terminates in a 3/8 inch (9.5 mm) barbed connector for standard cardiopulmonary bypass tubing. Each cannula is furnished with a mated obturator. Each obturator has a malleable wire. Each component is packaged inside a protective sheath in a single sterile, sealed pouch
Status
Terminated
Initiated
2017-09-28
Posted
not on file
Terminated
2018-10-17
Reason
A specific lot of the FlexFlowTM Venous Cannula, 200-200, is being recalled because it is sharper or pointier than expected.
Root cause
Process change control
Action
LivaNova sent an Urgent Medical Device Recall letter dated September 28 2017, to all affected consignees. The letter instructed customer to isolate products belonging to the lot involved. The level of the effectiveness check is Level A, where 100% of the consignees will be contacted. The customers were requested to send back any affected product and to complete the attached Customer Response Form by fax to 303-467-6502 or by email to USFSN@livanova.com. For questions customers were advised to contact Customer Service at 800-.650-2623.
Quantity
40 units
Distribution
Worldwide Distribution - US (nationwide) Internationally to France and Spain
Lot, serial or model codes
Lot: 181185; UDI: (01)10803622125812(240)200-200(17)200703(10)181185
510(k) numbers
["K052081"]
PMA numbers
not on file
Product code
DWF
Firm FEI
1718850
Firm city
Arvada
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register