FDA Medical Device Recalls (openFDA)
Zimmer, Inc.: NexGen Femoral Component (CR-FLEX GSF POR FEM, C-L; CR-FLEX GSF POR FEM, C-R; CR-FLEX GSF POR FEM, C-L MINUS; CR-FLEX GSF POR FEM, C-R MINUS; CR-FLEX GSF POR FEM, D-L; CR-FLEX GSF POR FEM, D-R CR-FLEX GSF POR FEM, D-L MINUS; CR-FLEX GSF POR FEM, D-R MINUS CR-FLEX GSF POR FEM, E-L; CR-FLEX GSF POR FEM, E-R; CR-FLEX GSF POR FEM, E-L MINUS; CR-FLEX GSF POR FEM, E-R MINUS; CR-FLEX GSF POR FEM, F-L ; CR-FLEX GSF POR FEM, F-R; CR-FLEX GSF POR FEM, F-L MINUS; CR-FLEX GSF POR FEM, F-R MINUS; CR-FLEX GSF POR FEM, G-L CR-FLEX GSF POR FEM, G-R; CR-FLEX GSF POR FEM, G-L MINUS CR-FLEX GSF POR FEM, G-R MINUS; NEXGEN GSF POROUS FEMORAL LPS FLEX SIZE C, LEFT; NEXGEN GSF POROUS FEMORAL LPS FLEX SIZE C, RIGHT; NEXGEN GSF POROUS FEMORAL LPS FLEX SIZE D, LEFT; NEXGEN GSF POROUS FEMORAL LPS FLEX SIZE D, RIGHT; LPS-FLEX GSF POR FEM, E-L; NEXGEN GSF POROUS FEMORAL LPS FLEX SIZE E, RIGHT; NEXGEN GSF POROUS FEMORAL LPS FLEX SIZE F, LEFT; NEXGEN GSF POROUS FEMORAL LPS FLEX SIZE F, RIGHT; NEXGEN GSF POROUS FEMORAL LPS FLEX SIZE G, LEFT; and NEXGEN GSF POROUS FEMORAL LPS FLEX SIZE G, RIGHT). Provides increased flexion capability.
- Recall number
- Z-0074-2014
- Recalling firm
- Zimmer, Inc.
- Product
- NexGen Femoral Component (CR-FLEX GSF POR FEM, C-L; CR-FLEX GSF POR FEM, C-R; CR-FLEX GSF POR FEM, C-L MINUS; CR-FLEX GSF POR FEM, C-R MINUS; CR-FLEX GSF POR FEM, D-L; CR-FLEX GSF POR FEM, D-R CR-FLEX GSF POR FEM, D-L MINUS; CR-FLEX GSF POR FEM, D-R MINUS CR-FLEX GSF POR FEM, E-L; CR-FLEX GSF POR FEM, E-R; CR-FLEX GSF POR FEM, E-L MINUS; CR-FLEX GSF POR FEM, E-R MINUS; CR-FLEX GSF POR FEM, F-L ; CR-FLEX GSF POR FEM, F-R; CR-FLEX GSF POR FEM, F-L MINUS; CR-FLEX GSF POR FEM, F-R MINUS; CR-FLEX GSF POR FEM, G-L CR-FLEX GSF POR FEM, G-R; CR-FLEX GSF POR FEM, G-L MINUS CR-FLEX GSF POR FEM, G-R MINUS; NEXGEN GSF POROUS FEMORAL LPS FLEX SIZE C, LEFT; NEXGEN GSF POROUS FEMORAL LPS FLEX SIZE C, RIGHT; NEXGEN GSF POROUS FEMORAL LPS FLEX SIZE D, LEFT; NEXGEN GSF POROUS FEMORAL LPS FLEX SIZE D, RIGHT; LPS-FLEX GSF POR FEM, E-L; NEXGEN GSF POROUS FEMORAL LPS FLEX SIZE E, RIGHT; NEXGEN GSF POROUS FEMORAL LPS FLEX SIZE F, LEFT; NEXGEN GSF POROUS FEMORAL LPS FLEX SIZE F, RIGHT; NEXGEN GSF POROUS FEMORAL LPS FLEX SIZE G, LEFT; and NEXGEN GSF POROUS FEMORAL LPS FLEX SIZE G, RIGHT). Provides increased flexion capability.
- Status
- Terminated
- Initiated
- 2013-08-30
- Posted
- 2013-10-25
- Terminated
- 2014-09-25
- Reason
- The low density polyethylene (LDPE) bag used to package implants adheres to the highly polished surface of the devices.
- Root cause
- Package design/selection
- Action
- August 30, 2013 Urgent Recall Notices were sent to Distributors, hospital risk managers and surgeons identifying the issue and directing consignees to make all affected personnel aware of the contents of the notification. The notice identified the products involved, the issue, possible adverse events associated with the issue and recommended that implants found with LDPE bag adhesion not use the product and return it to Zimmer sales representatives. Questions or concerns should be addressed to Zimmer customer call center at 1-877-946-2761 between the hours of 8 a.m. and 5 p.m. EST.
- Quantity
- 11694
- Distribution
- Worldwide Distribution: US (nationwide) and countries of: Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Germany, Guatemala, Honduras, Hong Kong, India, Israel, Jamaica, Japan, Korea, Malaysia, Mexico, New Zealand, Nicaragua, Panama, Peru, Singapore, Taiwan, Thailand, and Venezula.
- Lot, serial or model codes
- Part Number Range: 5752 and 5762.
- 510(k) numbers
- ["K060370"]
- PMA numbers
- not on file
- Product code
- JWH
- Firm FEI
- 1000220733
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28