Skip to the content

FDA Medical Device Recalls (openFDA)

CareFusion 203, Inc.: CareFusion EnVe Ventilator, Model 19250-001. Manufactured by: CareFusion, 22745 Savi Ranch, Yorba Linda, CA 92887 USA. EnVe is a portable critical care ventilator. It is suitable for a wide range of life support and critical care applications. The ventilator normally operates from external DC power. When the ventilator is portable, a Removable Battery Pack powers the unit. Devices are designed for use on patients who require respiratory support or mechanical ventilation and weigh a minimum of 5 kg (11 lbs).

Recall number
Z-0074-2012
Recalling firm
CareFusion 203, Inc.
Product
CareFusion EnVe Ventilator, Model 19250-001. Manufactured by: CareFusion, 22745 Savi Ranch, Yorba Linda, CA 92887 USA. EnVe is a portable critical care ventilator. It is suitable for a wide range of life support and critical care applications. The ventilator normally operates from external DC power. When the ventilator is portable, a Removable Battery Pack powers the unit. Devices are designed for use on patients who require respiratory support or mechanical ventilation and weigh a minimum of 5 kg (11 lbs).
Status
Terminated
Initiated
2011-09-12
Posted
2011-10-26
Terminated
2012-11-22
Reason
CareFusion has identified potential risks associated with the EnVe ventilator manufactured between December 2010 and May 2011. There are 3 issues that can result in ventilation delivery to the patient being interrupted. CareFusion will be contacting facilities to coordinate implementation of the necessary corrective actions.
Root cause
Device Design
Action
The firm, CareFusion, sent an "URGENT PRODUCT RECALL" letter dated September 12, 2011 to its customers.The letter identified affected product, described the 3 issues providing the description of the event, potential risk, and provided required action for Users. The letter also stated that CareFusion would be contacting the facility to coordinate the corrective action. If you have any questions and/or need support, please contact CareFusion Technical Support at 1-800-554-8933 from Mon. - Fri. 8:00am -5:00pm Pacific Time.
Quantity
128 units
Distribution
Nationwide distribution: USA including states of: CA, CO, FL, IL, MI, MN, MT, ND, NJ, NM, NY, OK, PA, TX, UT, WI, and WV.
Lot, serial or model codes
serial numbers: 45317, 45318, 45319, 45321, 45323, 45324, 45326, 45327, 45328, 45330, 45331, 45332, 45333, 45334, 45336, 45338, 45339, 45341, 45735, 45736, 45737, 45738, 45740, 45741, 45742, 45744, 45745, 45746, 45747, 45748, 45749, 45750, 45751, 45752, 45754, 45902, 45905, 45906, 45907, 45908, 45909, 45910, 46038, 46039, 46040, 46041, 46042, 46043, 46044, 46046, 46050, 46051, 46632, 46634, 46635, 46636, 46637, 46638, 46640, 46641, 46642, 46643, 46644, 46645, 47069, 47070, 47071, 47073, 47074, 47075, 47076, 47077, 47078, 47079, 47081, 47082, 47083, 47210, 47211, 47213, 47214, 47215, 47216, 47217, 47219, 47221, 47222, 47223, 47224, 47225, 47226, 47229, 47230, 47231, 47233, 47234, 47236, 47237, 47238, 47239, 47538, 47539, 47540, 47541, 47542, 47544, 47545, 47546, 47547, 47548, 47549, 47550, 47551, 47552, 47553, 47554, 47555, 47556, 47557, 47558, 47559, 47560, 47561, 47562, 47563, 47564, 47566, 47567.
510(k) numbers
["K070594"]
PMA numbers
not on file
Product code
CBK
Firm FEI
3002617992
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register