Medtronic Sofamor Danek USA Inc: Medtronic Sofamor Danek, TSRH SiLo 5.5 Spinal System, Sagittal Adjusting Screw Implants, Supplement Sets, REF: 8889245, Pedicle Screw Fixation System, Medtronic Sofamor Danek, Memphis, TN 38131
- Recall number
- Z-0074-2008
- Recalling firm
- Medtronic Sofamor Danek USA Inc
- Product
- Medtronic Sofamor Danek, TSRH SiLo 5.5 Spinal System, Sagittal Adjusting Screw Implants, Supplement Sets, REF: 8889245, Pedicle Screw Fixation System, Medtronic Sofamor Danek, Memphis, TN 38131
- Status
- Terminated
- Initiated
- 2007-08-24
- Posted
- 2007-12-13
- Terminated
- 2009-08-21
- Reason
- Disassembly: The movable cradle component could detach from the saddle under extreme manipulation of the associated rods prior to final tightening of the set screws.
- Root cause
- Device Design
- Action
- Medtronic sent letters, titled URGENT DEVICE RECALL, on 08/24/2007, to the Risk Managers of the hospital consignees explaining the potential for these sets to disassemble, and requesting the discontinuation of the use of and quarantine of the product. The product will be retrieved by the firm and replaced.
- Quantity
- 20 units
- Distribution
- Worldwide Distribution - USA, The Netherlands, Australia, and Canada.
- Lot, serial or model codes
- Lot Numbers: 1964441, 2633734, 14372033, 1813832, 2751495, 14673943, 14178935, 14013412, 14013714, 14013977, 14014230, 14014405, 14011615, 14011836, 15001233, 15001383, 13990466, 14103373, 14103376.
- 510(k) numbers
- ["K052144"]
- PMA numbers
- not on file
- Product code
- KWP
- Firm FEI
- 1000477302
- Firm city
- Memphis
- Firm state
- TN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28