Skip to the content

FDA Medical Device Recalls (openFDA)

Medtronic Sofamor Danek USA Inc: Medtronic Sofamor Danek, TSRH SiLo 5.5 Spinal System, Sagittal Adjusting Screw Implants, Supplement Sets, REF: 8889245, Pedicle Screw Fixation System, Medtronic Sofamor Danek, Memphis, TN 38131

Recall number
Z-0074-2008
Recalling firm
Medtronic Sofamor Danek USA Inc
Product
Medtronic Sofamor Danek, TSRH SiLo 5.5 Spinal System, Sagittal Adjusting Screw Implants, Supplement Sets, REF: 8889245, Pedicle Screw Fixation System, Medtronic Sofamor Danek, Memphis, TN 38131
Status
Terminated
Initiated
2007-08-24
Posted
2007-12-13
Terminated
2009-08-21
Reason
Disassembly: The movable cradle component could detach from the saddle under extreme manipulation of the associated rods prior to final tightening of the set screws.
Root cause
Device Design
Action
Medtronic sent letters, titled URGENT DEVICE RECALL, on 08/24/2007, to the Risk Managers of the hospital consignees explaining the potential for these sets to disassemble, and requesting the discontinuation of the use of and quarantine of the product. The product will be retrieved by the firm and replaced.
Quantity
20 units
Distribution
Worldwide Distribution - USA, The Netherlands, Australia, and Canada.
Lot, serial or model codes
Lot Numbers: 1964441, 2633734, 14372033, 1813832, 2751495, 14673943, 14178935, 14013412, 14013714, 14013977, 14014230, 14014405, 14011615, 14011836, 15001233, 15001383, 13990466, 14103373, 14103376.
510(k) numbers
["K052144"]
PMA numbers
not on file
Product code
KWP
Firm FEI
1000477302
Firm city
Memphis
Firm state
TN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register