Skip to the content

FDA Medical Device Recalls (openFDA)

Fujifilm Medical Systems U.S.A., Inc.: Fujifilm Synapse PACS, Software versions: 4.0.xxx, 4.1.xxx, version 4.2.xxx, version 4.3.xxx Medical imaging and information management system, SYNAPSE allows the archiving and distribution of image information from all modalities

Recall number
Z-0073-2017
Recalling firm
Fujifilm Medical Systems U.S.A., Inc.
Product
Fujifilm Synapse PACS, Software versions: 4.0.xxx, 4.1.xxx, version 4.2.xxx, version 4.3.xxx Medical imaging and information management system, SYNAPSE allows the archiving and distribution of image information from all modalities
Status
Terminated
Initiated
2016-09-15
Posted
not on file
Terminated
2019-02-20
Reason
Image data for a patients image may not be correct
Root cause
Software design
Action
The firm, FMSU, sent a "Medical Device Correction" letter dated September 15, 2016 to its customers. The letter describes the product, problem and actions to be taken. The firm will install a patch that corrects the issue (Synapse IPSS v1.3.1.0 Patch).Customers with Version 4.0 may upgrade to version 4.4 (will not require the patch) or upgrade to 4.1, 4.2, or 4.3 and install the patch. FMSU will install the patch on the server remotely if your site uses Active Monitoring. Your notification that the patch was installed will be your service ticket. For sites not using Active Monitoring, FMSU will schedule a time to install the patch on your server. Questions or concerns, contact the FUJIFILM Customer Support Center at 1-888-FUJIMED (1-800-385-4633) Option #1. International callers can access the FUJIFILM Customer Support Center by calling 1-203-602-3580.
Quantity
251 units
Distribution
US Nationwide Distribution.
Lot, serial or model codes
Software versions: 4.0.xxx, 4.1.xxx, version 4.2.xxx, version 4.3.xxx
510(k) numbers
["K112439"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
1000513161
Firm city
Stamford
Firm state
CT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register