FDA Medical Device Recalls (openFDA)
Acumed LLC: Biotrak PLLA Pin, Resorbable Fixation System, Biotrak 20 mm Pin; sterile pin packaged in sterile bag, enclosed within a box labeled Biotrak Resorbable Fixation System Biotrak 20 mm Pin REF 30170320-S Manufactured or Distributed by: ACUMED LLC 5885 NW Cornelius Pass Rd Hillsboro, OR 97124. The Acumed resorbable fixation devices are designed to provide fixation of specific fractures, osteotomies, and arthrodeses while they heal.
- Recall number
- Z-0073-2010
- Recalling firm
- Acumed LLC
- Product
- Biotrak PLLA Pin, Resorbable Fixation System, Biotrak 20 mm Pin; sterile pin packaged in sterile bag, enclosed within a box labeled Biotrak Resorbable Fixation System Biotrak 20 mm Pin REF 30170320-S Manufactured or Distributed by: ACUMED LLC 5885 NW Cornelius Pass Rd Hillsboro, OR 97124. The Acumed resorbable fixation devices are designed to provide fixation of specific fractures, osteotomies, and arthrodeses while they heal.
- Status
- Terminated
- Initiated
- 2009-08-21
- Posted
- 2009-10-23
- Terminated
- 2010-11-06
- Reason
- Labeling may contain 2 lot numbers and 2 model numbers. It was observed that the box top label (30170060-S, W67391) does not match the two side labels (3010320-S, W67385). The part and patient labeling inside the box match the side labels (30170320-S, W67385).
- Root cause
- Pending
- Action
- Urgent Notice: Device Recall Letters dated August 21, 2009 were sent to all consignees advising of the issue and and required action. Customers are to identify and quarantine the product having the affected lot code(s). Customers are to also return all product from the specified lot(s) to Acumed immediately. If product was further distributed, then those consignees should be notified of the recall and return of the product(s) should be requested. Questions should be directed to Acumed Customer Service at 1-888-627-9957.
- Quantity
- 125
- Distribution
- Worldwide Distribution -- USA and South Africa.
- Lot, serial or model codes
- Lot W67385, Model #30170320-S, expiration 8/2010 (two side labels correctly state 30170320-S, Lot W67385; top label may incorrectly state 30170060-S, lot W67391).
- 510(k) numbers
- ["K061763"]
- PMA numbers
- not on file
- Product code
- HWC
- Firm FEI
- 1000125930
- Firm city
- Hillsboro
- Firm state
- OR
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28