Spacelabs Healthcare, Ltd.: Arkon Anesthesia Delivery System, Model 99999, and software version 2.61 and upgrade kit 050-9043-00 Rev G. The Spacelabs Arkon Anesthesia Workstation is intended for use in the hospital environment and operating room. It may be used for the delivery of oxygen, air and nitrous oxide in a controlled manner to various patient breathing circuits with or without the use of mechanical ventilator, and may be used for the delivery of anesthetic vapor by use of a dismountable vaporizer. The device is intended for use only by a suitably qualified physician
- Recall number
- Z-0072-2018
- Recalling firm
- Spacelabs Healthcare, Ltd.
- Product
- Arkon Anesthesia Delivery System, Model 99999, and software version 2.61 and upgrade kit 050-9043-00 Rev G. The Spacelabs Arkon Anesthesia Workstation is intended for use in the hospital environment and operating room. It may be used for the delivery of oxygen, air and nitrous oxide in a controlled manner to various patient breathing circuits with or without the use of mechanical ventilator, and may be used for the delivery of anesthetic vapor by use of a dismountable vaporizer. The device is intended for use only by a suitably qualified physician
- Status
- Terminated
- Initiated
- 2017-10-11
- Posted
- 2017-11-21
- Terminated
- 2018-03-12
- Reason
- Arkon Anesthesia Workstation, with software version 2.61, experienced failure in mechanical ventilation, oxygen and anesthetic gas delivery, with concurrent failure of the display unit that resulted in a blank screen without audible or visible alarms.
- Root cause
- Software design
- Action
- Spacelabs Healthcare sent an Urgent Medical Device Correction letter dated October 24, 2017. The letter identified the affected product problem and actions to be taken. For questions call 1-800-522-7025 and select 4 for Technical Support.
- Quantity
- 110 units
- Distribution
- Nationwide Distribution to the states of NM, WY, FL, NC, MS, CO, CT, ME, and AL.
- Lot, serial or model codes
- UDI: 10841522100017
- 510(k) numbers
- ["K113051"]
- PMA numbers
- not on file
- Product code
- BSZ
- Firm FEI
- 3002807992
- Firm city
- Hertford
- Firm state
- not on file
- Firm country
- United Kingdom
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28