Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Howmedica Osteonics Corp.: Trident PSL HA Cluster Acetabular Shell; Arc Deposited HA Coated Cluster Screw Holes. Hip prosthesis component. Howmedica Osteonics Corp. 325 Corporate Drive; Mahwah, NJ 07430; A subsidiary of Stryker Corp. Sterile, Made in USA.

Recall number
Z-0072-2008
Recalling firm
Stryker Howmedica Osteonics Corp.
Product
Trident PSL HA Cluster Acetabular Shell; Arc Deposited HA Coated Cluster Screw Holes. Hip prosthesis component. Howmedica Osteonics Corp. 325 Corporate Drive; Mahwah, NJ 07430; A subsidiary of Stryker Corp. Sterile, Made in USA.
Status
Terminated
Initiated
2007-08-30
Posted
2007-10-25
Terminated
2008-09-05
Reason
Stryker Orthopaedics had identified that specific lots of Trident PSL Acetabular shells may have a dimensional discrepancy. The deviation regarding the difference in wall thickness will increase the gap between the shell and liner on one side and will decrease the gap between shell and liner on the opposing side, resulting in interference.
Root cause
Process control
Action
Recall notification letters were sent out on August 30, 2007 with return receipt.
Quantity
142 units
Distribution
Product was distributed nationwide and one unit to Brazil.
Lot, serial or model codes
Catalog # Description Lot code 542-11-46D TRIDENT PSL HA CLUSTER 46mm R3YMHD; 542-11-48D TRIDENT PSL HA CLUSTER 48mm 1PMMJD 542-11-48D TRIDENT PSL HA CLUSTER 48mm 663MJD 542-11-48D TRIDENT PSL HA CLUSTER 48mm 664MJD 542-11-50E TRIDENT PSL HA CLUSTER 50mm 1TEMJD 542-11-50E TRIDENT PSL HA CLUSTER 50mm 1TJMJD 542-11-50E TRIDENT PSL HA CLUSTER 50mm 2K0MJD 542-11-50E TRIDENT PSL HA CLUSTER 50mm 32JMJD 542-11-50E TRIDENT PSL HA CLUSTER 50mm 61YMJD 542-11-50E TRIDENT PSL HA CLUSTER 50mm 626MJD 542-11-50E TRIDENT PSL HA CLUSTER 50mm 70HMJD 542-11-50E TRIDENT PSL HA CLUSTER 50mm 71NMJD 542-11-50E TRIDENT PSL HA CLUSTER 50mm PYNMHD 542-11-50E TRIDENT PSL HA CLUSTER 50mm RHVMHD 542-11-50E TRIDENT PSL HA CLUSTER 50mm RHXMHD 542-11-52E TRIDENT PSL HA CLUSTER 52mm 8R8MJD 542-11-52E TRIDENT PSL HA CLUSTER 52mm 8RDMJD 542-11-52E TRIDENT PSL HA CLUSTER 52mm PYVMHD 542-11-52E TRIDENT PSL HA CLUSTER 52mm REKMHD 542-11-54F TRIDENT PSL HA CLUSTER 54mm 0YPMJD 542-11-54F TRIDENT PSL HA CLUSTER 54mm 3HAMJD 542-11-54F TRIDENT PSL HA CLUSTER 54mm 500MJD 542-11-54F TRIDENT PSL HA CLUSTER 54mm 76YMJD 542-11-54F TRIDENT PSL HA CLUSTER 54mm 7NWMJD 542-11-54F TRIDENT PSL HA CLUSTER 54mm 7Y3MJD 542-11-54F TRIDENT PSL HA CLUSTER 54mm 7Y4MJD 542-11-54F TRIDENT PSL HA CLUSTER 54mm 8HWMJD 542-11-54F TRIDENT PSL HA CLUSTER 54mm TK9MHD 542-11-54F TRIDENT PSL HA CLUSTER 54mm TKAMHD 542-11-54F TRIDENT PSL HA CLUSTER 54mm TKDMHD 542-11-54F TRIDENT PSL HA CLUSTER 54mm TKEMHD 542-11-54F TRIDENT PSL HA CLUSTER 54mm TKHMHD 542-11-56F TRIDENT PSL HA CLUSTER 56mm 0PJMJD 542-11-56F TRIDENT PSL HA CLUSTER 56mm 0PMMJD 542-11-56F TRIDENT PSL HA CLUSTER 56mm 81AMJD 542-11-56F TRIDENT PSL HA CLUSTER 56mm 81DMJD 542-11-56F TRIDENT PSL HA CLUSTER 56mm T6KMHD 542-11-58G TRIDENT PSL HA CLUSTER 58mm 518MJD 542-11-62H TRIDENT PSL HA CLUSTER 62mm 2PMMJD 542-11-62H TRIDENT PSL HA CLUSTER 62mm 7D2MJD 542-11-68I TRIDENT PSL HA CLUSTER 68mm 3AWMJD 542-11-72J TRIDENT PSL HA CLUSTER 72mm 1WNMJD
510(k) numbers
["K983382"]
PMA numbers
not on file
Product code
MEH
Firm FEI
3003070421
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register