Skip to the content

FDA Medical Device Recalls (openFDA)

American Contract Systems, Inc.: COVID Test Kit Nonsterile

Recall number
Z-0071-2023
Recalling firm
American Contract Systems, Inc.
Product
COVID Test Kit Nonsterile
Status
Open, Classified
Initiated
2022-08-12
Posted
2022-10-08
Terminated
not on file
Reason
Kit incorrectly labeled for use with COVID testing and should not be used with any SARS-Co-V2 device.
Root cause
No Marketing Application
Action
A customer letter dated 08/12/2022 was sent on or around 08/15/2022 to the single impacted consignee. The customer is instructed that the product cannot be used in association with SARS-CoV-2 or COVID-19 testing but instead only for general use of specimen collection. The product is to be relabeled in accordance with the recall communication letter. A Field Action Response Form is to be completed and returned to angela.bunn@owens-minor.com. A second customer letter dated September 8, 2022 was sent on or around the same day to notify the consignee that 2 lots listed in this event cannot be relabeled, as they are subject to destruction from a previous recall event. Those two lots are Product #AHCV18A, lots 980221 and 974221. The consignee is requested to confirm destruction of the lots.
Quantity
39,575 kits
Distribution
Distribution to one customer in PA only
Lot, serial or model codes
1) UDI-DI: Product Code: AHCV18A, Lots 980221 (exp. 01/20/2023) and 974221 (exp. 01/25/2023) 2) UDI-DI: Product Code: AHCV18B, Lots 868221 (exp. 05/31/2023), 842221 (exp. 05/31/2023)
510(k) numbers
not on file
PMA numbers
not on file
Product code
JSM
Firm FEI
3008908305
Firm city
Bloomington
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register