Skip to the content

FDA Medical Device Recalls (openFDA)

Cardiocommand Inc.: ***REF #200***TAPSCOPE 550. Each pouch contains one (1) 18 French esophageal stethoscope with bipolar pacing and recording electrodes with a YSI 400 series male connector thermistor. LOT xxxxxxx 400095-0200-C 100902. www.cardiocommand.com. ***REF #200-0010***TAPSCOPE 550 CONTAINS: Ten (10) 18 French esophageal stethoscopes with bipolar pacing and recording electrodes with YSI 400 series male connector thermistors. NON-STERILE. SINGLE USE ONLY. DO NOT REUSE. CAUTION: Federal (U.S.A.) law restricts this device to sale, distribution and use by or on the order of a licensed physician. Manufacturer: CARDIOCOMMAND, INC. Tampa, FL USA. (800) 231-6370 (813) 289-5555. www.cardiocommand.com. LOT xxxxxxx 400155-B 091101. ***REF #200-0025***TAPSCOPE 550 Contains: Twenty five (25) 18 French esophageal stethoscopes with bipolar pacing and recording electrodes with YSI 400 series male connector thermistors. ***LOT xxxxxxx***FOR SINGLE USE ONLY. DO NOT REUSE. NON-STERILE. CAUTION: Federal (U.S.A.) law restricts this device to sale, distribution and use by or on the order of a licensed physician. Read Model 2A operator's manual prior to initiating pacing procedures. Manufacturer: Tampa, FL USA 33607. (800) 231-6370 (813) 289-5555. www.cardiocommand.com. 400215-A 100902.

Recall number
Z-0071-2012
Recalling firm
Cardiocommand Inc.
Product
***REF #200***TAPSCOPE 550. Each pouch contains one (1) 18 French esophageal stethoscope with bipolar pacing and recording electrodes with a YSI 400 series male connector thermistor. LOT xxxxxxx 400095-0200-C 100902. www.cardiocommand.com. ***REF #200-0010***TAPSCOPE 550 CONTAINS: Ten (10) 18 French esophageal stethoscopes with bipolar pacing and recording electrodes with YSI 400 series male connector thermistors. NON-STERILE. SINGLE USE ONLY. DO NOT REUSE. CAUTION: Federal (U.S.A.) law restricts this device to sale, distribution and use by or on the order of a licensed physician. Manufacturer: CARDIOCOMMAND, INC. Tampa, FL USA. (800) 231-6370 (813) 289-5555. www.cardiocommand.com. LOT xxxxxxx 400155-B 091101. ***REF #200-0025***TAPSCOPE 550 Contains: Twenty five (25) 18 French esophageal stethoscopes with bipolar pacing and recording electrodes with YSI 400 series male connector thermistors. ***LOT xxxxxxx***FOR SINGLE USE ONLY. DO NOT REUSE. NON-STERILE. CAUTION: Federal (U.S.A.) law restricts this device to sale, distribution and use by or on the order of a licensed physician. Read Model 2A operator's manual prior to initiating pacing procedures. Manufacturer: Tampa, FL USA 33607. (800) 231-6370 (813) 289-5555. www.cardiocommand.com. 400215-A 100902.
Status
Terminated
Initiated
2010-10-25
Posted
2011-10-19
Terminated
2012-08-28
Reason
CardioCommand, Inc. is recalling their TAPSCOPE 550 esophageal pacing/recording catheters because there maybe electrical discontinuity, which can result in the inability to reliably pace and/or record ECG.
Root cause
Nonconforming Material/Component
Action
On October 25, 2010, an Urgent Medical Device Recall Notice was sent by certified mail to all facilities advising them of the recall. The letter identified the affected product and reason for the recall. Users were asked to discontinue use and return the affected product to the firm. If all product was already used, then this should be indicated on the form provided. The firm will send replacement product in quantities equal to the returned product. The attached form should be completed and returned via fax. Customers were instructed to report any suspected incidents related to this issue to CardioCommand Customer Service at 1-800-231-6370. Also questions or comments should be directed to CardioCommand Customer Service.
Quantity
567 total
Distribution
Nationwide Distribution -- GA, OR, TX, CO, KY, MT, MA, AL, CA, FL & TN.
Lot, serial or model codes
Lot Numbers: 1026504, 1023803, 1022901, 1019007, and 1020106.
510(k) numbers
not on file
PMA numbers
["P860008"]
Product code
LPA
Firm FEI
1927197
Firm city
Tampa
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register