FDA Medical Device Recalls (openFDA)
Cardiocommand Inc.: ***REF #200***TAPSCOPE 550. Each pouch contains one (1) 18 French esophageal stethoscope with bipolar pacing and recording electrodes with a YSI 400 series male connector thermistor. LOT xxxxxxx 400095-0200-C 100902. www.cardiocommand.com. ***REF #200-0010***TAPSCOPE 550 CONTAINS: Ten (10) 18 French esophageal stethoscopes with bipolar pacing and recording electrodes with YSI 400 series male connector thermistors. NON-STERILE. SINGLE USE ONLY. DO NOT REUSE. CAUTION: Federal (U.S.A.) law restricts this device to sale, distribution and use by or on the order of a licensed physician. Manufacturer: CARDIOCOMMAND, INC. Tampa, FL USA. (800) 231-6370 (813) 289-5555. www.cardiocommand.com. LOT xxxxxxx 400155-B 091101. ***REF #200-0025***TAPSCOPE 550 Contains: Twenty five (25) 18 French esophageal stethoscopes with bipolar pacing and recording electrodes with YSI 400 series male connector thermistors. ***LOT xxxxxxx***FOR SINGLE USE ONLY. DO NOT REUSE. NON-STERILE. CAUTION: Federal (U.S.A.) law restricts this device to sale, distribution and use by or on the order of a licensed physician. Read Model 2A operator's manual prior to initiating pacing procedures. Manufacturer: Tampa, FL USA 33607. (800) 231-6370 (813) 289-5555. www.cardiocommand.com. 400215-A 100902.
- Recall number
- Z-0071-2012
- Recalling firm
- Cardiocommand Inc.
- Product
- ***REF #200***TAPSCOPE 550. Each pouch contains one (1) 18 French esophageal stethoscope with bipolar pacing and recording electrodes with a YSI 400 series male connector thermistor. LOT xxxxxxx 400095-0200-C 100902. www.cardiocommand.com. ***REF #200-0010***TAPSCOPE 550 CONTAINS: Ten (10) 18 French esophageal stethoscopes with bipolar pacing and recording electrodes with YSI 400 series male connector thermistors. NON-STERILE. SINGLE USE ONLY. DO NOT REUSE. CAUTION: Federal (U.S.A.) law restricts this device to sale, distribution and use by or on the order of a licensed physician. Manufacturer: CARDIOCOMMAND, INC. Tampa, FL USA. (800) 231-6370 (813) 289-5555. www.cardiocommand.com. LOT xxxxxxx 400155-B 091101. ***REF #200-0025***TAPSCOPE 550 Contains: Twenty five (25) 18 French esophageal stethoscopes with bipolar pacing and recording electrodes with YSI 400 series male connector thermistors. ***LOT xxxxxxx***FOR SINGLE USE ONLY. DO NOT REUSE. NON-STERILE. CAUTION: Federal (U.S.A.) law restricts this device to sale, distribution and use by or on the order of a licensed physician. Read Model 2A operator's manual prior to initiating pacing procedures. Manufacturer: Tampa, FL USA 33607. (800) 231-6370 (813) 289-5555. www.cardiocommand.com. 400215-A 100902.
- Status
- Terminated
- Initiated
- 2010-10-25
- Posted
- 2011-10-19
- Terminated
- 2012-08-28
- Reason
- CardioCommand, Inc. is recalling their TAPSCOPE 550 esophageal pacing/recording catheters because there maybe electrical discontinuity, which can result in the inability to reliably pace and/or record ECG.
- Root cause
- Nonconforming Material/Component
- Action
- On October 25, 2010, an Urgent Medical Device Recall Notice was sent by certified mail to all facilities advising them of the recall. The letter identified the affected product and reason for the recall. Users were asked to discontinue use and return the affected product to the firm. If all product was already used, then this should be indicated on the form provided. The firm will send replacement product in quantities equal to the returned product. The attached form should be completed and returned via fax. Customers were instructed to report any suspected incidents related to this issue to CardioCommand Customer Service at 1-800-231-6370. Also questions or comments should be directed to CardioCommand Customer Service.
- Quantity
- 567 total
- Distribution
- Nationwide Distribution -- GA, OR, TX, CO, KY, MT, MA, AL, CA, FL & TN.
- Lot, serial or model codes
- Lot Numbers: 1026504, 1023803, 1022901, 1019007, and 1020106.
- 510(k) numbers
- not on file
- PMA numbers
- ["P860008"]
- Product code
- LPA
- Firm FEI
- 1927197
- Firm city
- Tampa
- Firm state
- FL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28