Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Sienet Sky, picture archiving and communications system. Model number 8463163

Recall number
Z-0071-2007
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Sienet Sky, picture archiving and communications system. Model number 8463163
Status
Terminated
Initiated
2006-09-05
Posted
2006-10-25
Terminated
2008-09-17
Reason
Potential issue with annotation function. Annotations, shown as overlays, may change font size when the image is saved. Text may appear shifted, or lines appear thicker, and their position may not be as intended.
Root cause
Other
Action
The recalling firm has issued a Customer Safety Advisory via certified mail to affected customers per Update Instructions HS096/06/S and HS097/06/S. The letter, dated September 5, 2006, informs customers of this issue and provides recommendations to avoid its occurrence.
Quantity
1818 units total, for all three products
Distribution
The product was shipped to medical facilities nationwide.
Lot, serial or model codes
Model number 8463163, all versions.
510(k) numbers
["K935694"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register