Medtronic Navigation, Inc.-Littleton: Medtronic Navigation O-arm O2 Surgical Imaging System Product Catalog Number: BI-700-02000 Product Usage: The O-arm Surgical Imaging System is a multi-dimensional surgical imaging platform that is designed for use in spine, orthopaedic, and trauma-related surgeries. It provides real-time, intra-operative imaging of a patients anatomy with high quality images and a large field-of-view in both two and three dimensions

Recall number
Z-0070-2017
Recalling firm
Medtronic Navigation, Inc.-Littleton
Product
Medtronic Navigation O-arm O2 Surgical Imaging System Product Catalog Number: BI-700-02000 Product Usage: The O-arm Surgical Imaging System is a multi-dimensional surgical imaging platform that is designed for use in spine, orthopaedic, and trauma-related surgeries. It provides real-time, intra-operative imaging of a patients anatomy with high quality images and a large field-of-view in both two and three dimensions
Status
Terminated
Initiated
2016-09-22
Posted
2016-10-11
Terminated
2017-05-12
Reason
O-arm O2 Surgical Imaging System Spatial calibration may be erroneous in Stealth Station navigated images
Root cause
Software design
Action
Medtronic Navigation contacted customers via telephone beginning September 23, 2016. Customers were informed of the affected product, problem and actions to be taken. Medtronic requests that users immediately discontinue the use of the O-Arm O2 Surgical Imaging system for automatic registration in Stealth Station navigated surgical procedures until further notice. Medtronic Service personnel have been scheduled to begin the necessary inspections and re-calibration activities on customer systems beginning on September 23, 2016.
Quantity
14 units
Distribution
Worldwide Distribution - US Nationwide in the states of CO, KA, NE, MI, MO, OH, WA and country of Switzerland
Lot, serial or model codes
Software revisions 4.0.0, 4.0.1 and 4.0.2. System Serial Numbers: C0930  C0981  C0986  C1113 C1069 C1050 C1208 C1043 C1144 C1163 C1145 C1162 C1002 C1074
510(k) numbers
["K151000"]
PMA numbers
not on file
Product code
OWB
Firm FEI
3004785967
Firm city
Littleton
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register